Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support...
Boston Scientific Corporation began a recall of Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Application on . FDA classified it as Class II and gives the reason as "There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0319-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Application
- Product code
- LWS Implantable Cardioverter Defibrillator (Non-Crt)
- Reason for recall
- There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Boston Scientific sales representatives will hand-deliver a removal letter to all customer consignees with one of the eight affected Model 3300 LATITUDE programmer. BSC sales representatives will maintain control of these devices. In order to continue to provide the benefits that this software support application provides for those patients implanted with an ALTRUA/INSIGNIA I pacemaker, BSC sales representatives will continue to use Model 3300 LATITUDE programmers installed with the Model 3892 v1.04 ALTRUA / INSIGNIA I / NEXUS I Software Support Application in the field. The risks associated with this unintended behavior can be avoided when specific remedial actions are adhered to. These remedial actions will be adhered to until a software correction is approved and installed on these programmers. BSC will confirm via email from sales representatives that 100% of the removal letters were delivered. BSC will post the removal letter on the BSC website within the Product Performance Resource Center located @ www.bostonscientific.com/en-US/pprc/productadvisories.html.
- Quantity in commerce
- 8 devices
- Distribution
- Distributed in US - AR, FL, IL, and TX
- Product codes and lots
- Model 3892 v1.04 is installed on Model 3300 LATITUDE Programming Systems with the following serial numbers: 009494 012689 010356 008206 008124 007983 004813 010751
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0319-2021 | Class II | Terminated | Voluntary: Firm initiated |