Z-0319-2021Class II

Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support...

Boston Scientific Corporation / initiated 2020-09-22 / Terminated / root cause: software

Boston Scientific Corporation began a recall of Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Application on . FDA classified it as Class II and gives the reason as "There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0319-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Application
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line.
Root cause tag
software
FDA root cause
Software design
Action
Boston Scientific sales representatives will hand-deliver a removal letter to all customer consignees with one of the eight affected Model 3300 LATITUDE programmer. BSC sales representatives will maintain control of these devices. In order to continue to provide the benefits that this software support application provides for those patients implanted with an ALTRUA/INSIGNIA I pacemaker, BSC sales representatives will continue to use Model 3300 LATITUDE programmers installed with the Model 3892 v1.04 ALTRUA / INSIGNIA I / NEXUS I Software Support Application in the field. The risks associated with this unintended behavior can be avoided when specific remedial actions are adhered to. These remedial actions will be adhered to until a software correction is approved and installed on these programmers. BSC will confirm via email from sales representatives that 100% of the removal letters were delivered. BSC will post the removal letter on the BSC website within the Product Performance Resource Center located @ www.bostonscientific.com/en-US/pprc/productadvisories.html.
Quantity in commerce
8 devices
Distribution
Distributed in US - AR, FL, IL, and TX
Product codes and lots
Model 3892 v1.04 is installed on Model 3300 LATITUDE Programming Systems with the following serial numbers: 009494 012689 010356 008206 008124 007983 004813 010751

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0319-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0319-2021Class IITerminatedVoluntary: Firm initiated