Z-0287-2021Class II
Visual-ICE MRI Cryoablation System Mobile Connection Panel (Visual-ICE MRI MCP) -
Boston Scientific Corporation / initiated 2020-09-22 / Terminated / root cause: manufacturing process
Boston Scientific Corporation began a recall of Visual-ICE MRI Cryoablation System Mobile Connection Panel (Visual-ICE MRI MCP) - Product Usage: The system is computer-controlled with a touch screen user... on . FDA classified it as Class II and gives the reason as "Incorrectly installed base". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0287-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Visual-ICE MRI Cryoablation System Mobile Connection Panel (Visual-ICE MRI MCP) - Product Usage: The system is computer-controlled with a touch screen user interface that allows the user to control and monitor the procedure.
- Product code
- GEH Unit, Cryosurgical, Accessories
- Reason for recall
- Incorrectly installed base.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- On 9/22/2020 the firm sent its consignee a letter. Accounts are not asked to return the affected product to BSC. The customer notification letter will notify the affected customer that the Visual-ICE MRI MCP unit will be corrected, and accounts are asked to complete an Acknowledgment Form indicating they have received the customer notification and are scheduling the field correction activities.
- Quantity in commerce
- 1 unit
- Distribution
- US Nationwide distribution including in the state of Rochester, MN.
- Product codes and lots
- Serial Number: VM031