VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Control -
Ortho Clinical Diagnostics Inc began a recall of VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Control - Product Usage: For use in monitoring the performance of the VITROS Immunodiagnostic and... on . FDA classified it as Class II and gives the reason as "Elevated sample results signal/cutoff (s/c) values when using eight (8) lots of VITROS immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator and decreased signal in one (1) lot of VITROS Immunodiagnostic Products...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0451-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Control - Product Usage: For use in monitoring the performance of the VITROS Immunodiagnostic and Integrated Systems when used for the determination of IgG antibodies to SARS-CoV-2. Product Code (Unique Identifier): 6199921 (10758750033355)
- Product code
- QMN Covid-19 Multi-Analyte Antigen Device
- Reason for recall
- Elevated sample results signal/cutoff (s/c) values when using eight (8) lots of VITROS immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator and decreased signal in one (1) lot of VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Control
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Ortho Clinical Diagnostics issued Urgent Product Correction Notice on 30 Sept 2020 via(CL2020-246) of the issue and instructed as follows: If they are unable to calibrate (now or in the future) using an affected lot of VITROS Anti-SARS-CoV-2 IgG Calibrator Ortho will replace their remaining inventory as indicated on their Confirmation of Receipt form. NOTE: If they have a successful calibration, and quality control results are acceptable, the laboratory may continue using the affected lots for qualitative detection of IgG antibodies to SARS-CoV-2. If they are using the affected product to qualify units as high titer for CPT, consult with your medical director to determine if they should continue testing while waiting for their replacement product. Ortho will replace or credit their remaining inventory as indicated on their Confirmation of Receipt form. Post this notification by each system that processes the affected lots of VITROS Anti-SARS-CoV-2 IgG Calibrator or Control. Complete the enclosed Confirmation of Receipt form no later than October 9, 2020. Indicate the quantity of their remaining inventory to be replaced or credited. Please forward this notification if the product was distributed outside of their facility
- Quantity in commerce
- 691 units (OUS)
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Australia, Brazil, Chile, China, Colombia, India, Japan, Singapore 609917, United Kingdom, Italy.
- Product codes and lots
- Lot Number/ Exp. Date: 0110 25-NOV-20
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0450-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0451-2021 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
