Each KWIK-STIK unit contains a lyophilized microorganism pellet, an ampoule of hydrating fluid, and an...
Microbiologics Inc began a recall of Each KWIK-STIK unit contains a lyophilized microorganism pellet, an ampoule of hydrating fluid, and an inoculating swab. Each device is sealed within a... on . FDA classified it as Class II and gives the reason as "KWIK-STIK 6 Pack Klebsiella pneumoniae derived from NCTC 13443 (Catalog# 01145K) should contain the NDM-1 positive strain, but was confirmed to be NDM-1 negative". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0303-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Microbiologics, Inc.
- Product
- Each KWIK-STIK unit contains a lyophilized microorganism pellet, an ampoule of hydrating fluid, and an inoculating swab. Each device is sealed within a laminated pouch that contains a desiccant to prevent adverse moisture accumulation. KWIK-STIK microorganisms are 3 passages or fewer from the reference culture and are guaranteed to recover when processed using the recommended media and incubation requirements. Available in packs of 2 or 6.
- Product code
- JTR Kit, Quality Control For Culture Media
- Reason for recall
- KWIK-STIK 6 Pack Klebsiella pneumoniae derived from NCTC 13443 (Catalog# 01145K) should contain the NDM-1 positive strain, but was confirmed to be NDM-1 negative.
- FDA root cause
- Under Investigation by firm
- Action
- The recalling firm is sending consignees recall notification letters beginning 10/08/2020 via email. The correction communication was first extended to the first level of the distribution change (those who received product directly from Microbiologics). Distributors outside of the US will be contacted via email, and if follow up is necessary, the recalling firm's international sales team will work with each distributor to ensure this communication is extended to their end users. The recalling firm is working with its European Authorized Representative to coordinate this field safety corrective action in the EU.
- Quantity in commerce
- 1 unit
- Distribution
- Distribution in US, Brazil, Canada, Costa Rica, France, Germany, Ireland, Italy, Japan, New Zealand, Sweden, and United Kingdom.
- Product codes and lots
- Catalog 01145K, Lot 1145-06-5 UDI: 30845357036811
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0302-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0303-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0304-2021 | Class II | Terminated | Voluntary: Firm initiated |