Streamline TL Navigation System, Catalog Nos. 02-C-TAP-55-MDN 02-C-TAP-60-L-MDN 02-C-TAP-65-L-MDN...
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) began a recall of Streamline TL Navigation System, Catalog Nos. 02-C-TAP-55-MDN 02-C-TAP-60-L-MDN 02-C-TAP-65-L-MDN 02-C-TAP-65-MDN 02-C-TAP-75-L-MDN 02-C-TAP-85-L-MDN... on . FDA classified it as Class II and gives the reason as "The Streamline Navigation System taps have experienced binding with other devices". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0300-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)
- Product
- Streamline TL Navigation System, Catalog Nos. 02-C-TAP-55-MDN 02-C-TAP-60-L-MDN 02-C-TAP-65-L-MDN 02-C-TAP-65-MDN 02-C-TAP-75-L-MDN 02-C-TAP-85-L-MDN 02-C-TAP-85-MDN 02-C-TAP-95-L-MDN 02-RS-INS-C-MDN 02-RS-INSERTER-MDN 02-SCREWINS-CL-MDN 02-SCREWINS-C-MDN 02-S-FINDER-L-MDN 02-SFINDER-MDN 02-S-FINDER-T-MDN 02-TAP-45-MDN 02-TAP-55-MDN 02-TAP-60-L-MDN 02-TAP-65-L-MDN 02-TAP-65-MDN 02-TAP-75-L-MDN 02-TAP-75-MDN 02-TAP-85-L-MDN 02-TAP-85-MDN 02-TAP-95-L-MDN 02-MODHANDLE-MDN Product Usage: intended to be used during the preparation and placement of screws from the Streamline TL and MIS Systems and preparation of screws from the Streamline OCT System during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures.
- Product code
- OLO Orthopedic Stereotaxic Instrument
- Reason for recall
- The Streamline Navigation System taps have experienced binding with other devices.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On September 24, 2020, the firm distributed the Urgent Medical Devices Recall letter to customers. Customers were instructed to remove affected product from all user sites and inventory locations. These instrument systems are to be returned to RTI Surgical along with a completed copy of the response form. Accounts who have further distributed the affected products are also asked to immediately notify their customers by providing them a copy of the recall letter and the response form for completion when returning the affected devices to RTI Surgical.
- Quantity in commerce
- 420
- Distribution
- US Nationwide distribution including in the state of Michigan.
- Product codes and lots
- Lots 372816 377619 372817 372813 372927 373852 377616 377620 372926 372925 372923 372924 366798 367040 366799 370212 371729 377621 371730 373437 371731 372815 377622 373845 377623 369372 370853 377624
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0299-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0300-2021 | Class II | Terminated | Voluntary: Firm initiated |