Z-0322-2021Class II

Streamline OCT Navigated Instruments, Catalog numbers 26-DRILL-MDN 26-FTAP-30-MDN 26-FTAP-35-MDN...

RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) / initiated 2020-09-24 / Open, Classified / root cause: design

RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) began a recall of Streamline OCT Navigated Instruments, Catalog numbers 26-DRILL-MDN 26-FTAP-30-MDN 26-FTAP-35-MDN 26-FTAP-40-MDN 26-FTAP-40-MDN 26-FTAP-45-MDN 26-S-FINDER-MDN... on . FDA classified it as Class II and gives the reason as "System does not include a drill guide". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0322-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)
Product
Streamline OCT Navigated Instruments, Catalog numbers 26-DRILL-MDN 26-FTAP-30-MDN 26-FTAP-35-MDN 26-FTAP-40-MDN 26-FTAP-40-MDN 26-FTAP-45-MDN 26-S-FINDER-MDN Components of a surgical navigation device.
Product code
OLO Orthopedic Stereotaxic Instrument
Reason for recall
System does not include a drill guide.
Root cause tag
design
FDA root cause
Device Design
Action
On September 24, 2020, the firm sent an Urgent Medical Devices Recall letter to customers to inform them of the product issue. Customers were instructed to remove affected product from all user sites and inventory locations. Customers are to return these instrument systems to RTI Surgical using RGA number 63000897 along with a completed copy of the Acknowledgement and Receipt Form. If the product has been further distributed, please immediately notify your customers by providing them a copy of the recall letter and the response form for completion.
Quantity in commerce
76 units
Distribution
Distributed to one distributor in Michigan.
Product codes and lots
Lots  370855 373847 373848 373849 377629 373850 373846

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0322-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0322-2021Class IIOngoingVoluntary: Firm initiated