Z-0458-2021Class III
Olympus Leak Test Adapter used with Acu-sInQ Complete Endoscope Cleaning Aid Systems-intended to assist...
STERIS Corporation / initiated 2020-10-07 / Terminated / root cause: design
Steris Corporation began a recall of Olympus Leak Test Adapter used with Acu-sInQ Complete Endoscope Cleaning Aid Systems-intended to assist manual flexible endoscope cleaning Part number... on . FDA classified it as Class III and gives the reason as "Potential for cleaning solution intrusion into endoscope". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0458-2021
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- STERIS Corporation
- Product
- Olympus Leak Test Adapter used with Acu-sInQ Complete Endoscope Cleaning Aid Systems-intended to assist manual flexible endoscope cleaning Part number: 0600818OKN
- Product code
- FEB Accessories, Cleaning, For Endoscope
- Reason for recall
- Potential for cleaning solution intrusion into endoscope
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Steris issued Urgent Medical Device Recall issued letter on 10/7/20 stating reason for recall, health risk and action to take: Please immediately inspect your facility for affected Leak Test Adapters that connect to Olympus endoscopes. The affected Leak Test Adapters have a white plastic valve connector. If you have this Leak Test Adapter, please disconnect from your Acu-sInQ Complete, destroy, and replace with the new Leak Test Adapter included with this Customer Notification Letter. 2. Please complete the Medical Device Recall Response Form included with this Customer Notification Letter. Your STERIS Sales Representative can assist you should you have any questions while completing the form, or when replacing the Leak Test Adapter. 3. Return the completed Medical Device Recall Response Form to STERIS via one of the following three methods: a. Visit https://www.steris.com/medical-device-recall-response-form and complete the form electronically; b. Scan and email the completed form to [email protected]; or c. Fax the completed form to (440) 392-8963. We apologize for any inconvenience this matter may cause, and as always, STERIS is dedicated to supporting our products and valued Customers. If you have questions regarding this matter, please contact Katilin Hawranko, Product Manager at (440) 392-7431, STERIS Customer Service at 1-800-548-4873, or your local STERIS Representative.
- Quantity in commerce
- 293 units:248 units (USA); 45 units OUS
- Distribution
- Nationwide Foreign: Canada, Italy, Kuwait, United Kingdom, United Arab Emirates
- Product codes and lots
- Distributed from May 23, 2017 to June 30, 2020
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0458-2021 | Class III | Terminated | Voluntary: Firm initiated |