Z-0457-2021Class II

Osteosynthesis Compression Staple EasyClip

Stryker Corp. / initiated 2020-09-25 / Open, Classified

Stryker GmbH began a recall of Osteosynthesis Compression Staple EasyClip on . FDA classified it as Class II and gives the reason as "The device has the potential to release nickel above the acceptable margin of safety in pediatric patients who weigh less than 20 kg (44.09 lbs) when 2 or more implants are used". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0457-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Stryker Corp.
Product
Osteosynthesis Compression Staple EasyClip
Product code
JDR Staple, Fixation, Bone
Reason for recall
The device has the potential to release nickel above the acceptable margin of safety in pediatric patients who weigh less than 20 kg (44.09 lbs) when 2 or more implants are used.
FDA root cause
Under Investigation by firm
Action
All affected customers were mailed an Urgent Medical Device Correction (UMDC) letter on 09/25/2020 that requested them to take the following actions: 1. Please inform users of this Urgent Medical Device Correction (UMDC) and forward this notice to all individuals who need to be made aware of the recall. 2. Under 21 CFR 803, manufacturers are also required to report any serious injuries where a product has contributed or may have contributed to the event. Please keep Stryker informed of any adverse events associated with this product by emailing [email protected]. 3. A response is required, even though you may not have any physical inventory on site anymore. 4. Hospitals/Branches/Agencies: Complete and sign the enclosed Urgent Medical Device Correction Business Reply Form and fax a copy to 1-888-240-4294 or email to [email protected]. 5. Branches/Agencies Only: Please forward this Urgent Medical Device Correction to the individuals or organizations who have consigned product, if applicable. Please assist us in meeting our regulatory obligation by faxing back the attached Urgent Medical Device Correction Business Reply Form within 5 days. We regret any inconvenience this action may cause you and if you have any questions or concerns after reviewing the letter, please contact Customer Service at (201) 831-5000. For questions pertaining to the recall, email [email protected]
Quantity in commerce
121,517 EasyClip devices and 9,561 EasyClip Xpress devices distributed nationwide in US.
Distribution
The devices were distributed nationwide throughout the US, and to countries OUS. Easy Clip devices: As of 09/04/2020, 251,566 EasyClip devices have been manufactured. 121,517 of those devices were distributed to the US, and 102,529 were distributed in countries OUS. 27,520 devices were left in the firm's inventory. Easy Clip Xpress devices: As of 09/04/2020, 37,371 EasyClip Xpress devices have been manufactured. 9,561 of those devices were distributed to the US, leaving 27,810 in the firm's inventory. This device is only distributed in the United States.
Product codes and lots
All lots of the Osteosynthesis Compression Staple EasyClip and EasyClip Xpress are affected. Exp. Date: 5 years from date of manufacturing  EasyClip and EasyClip Xpress staples are available in a variety of sizes identified by product and catalog numbers. See below:  Product Number Product Description  EZ15-12-12 Osteosynthesis Compression Staple EasyClip 15x12x12mm EZ15-15-15 Osteosynthesis Compression Staple EasyClip 15x15x15mm  EZ18-12-12 Osteosynthesis Compression Staple EasyClip 18x12x12mm EZ18-15-13 Osteosynthesis Compression Staple EasyClip 18x15x13mm  EZ18-14-14 Osteosynthesis Compression Staple EasyClip 18x14x14mm EZ18-17-15 Osteosynthesis Compression Staple EasyClip 18x17x15mm  EZ18-19-17 Osteosynthesis Compression Staple EasyClip 18x19x17mm EZ20-16-16 Osteosynthesis Compression Staple EasyClip 20x16x16mm EZ20-20-20 Osteosynthesis Compression Staple EasyClip 20x20x20mm EZ25-22-22 Osteosynthesis Compression Staple EasyClip 25x22x22mm EZB10-15-13 Osteosynthesis Compression Staple EasyClip 10x15x13mm EZB10-17-15 Osteosynthesis Compression Staple EasyClip 10x17x15mm EZB10-19-16 Osteosynthesis Compression Staple EasyClip 10x19x16mm EZB10-20-17 Osteosynthesis Compression Staple EasyClip 10x20x17mm EZB12-15-13 Osteosynthesis Compression Staple EasyClip 12x15x13mm EZB10-22-20 Osteosynthesis Compression Staple EasyClip 10x22x20mm  EZB12-17-15 Osteosynthesis Compression Staple EasyClip 12x17x15mm EZB12-19-16 Osteosynthesis Compression Staple EasyClip 12x19x16mm  EZM08-08-08 Osteosynthesis Compression Staple EasyClip 8x8x8mm EZM10-10-10 Osteosynthesis Compression Staple EasyClip 10x10x10mm EZM12-10-10 Osteosynthesis Compression Staple EasyClip 10x10x10mm EZM12-14-14 Osteosynthesis Compression Staple EasyClip 12x14x14mm EZM12-16-16 Osteosynthesis Compression Staple EasyClip 12x16x16mm   Catalog # Product EZBXP10-15-13 EasyClip Xpress 10x15x13 EZBXP12-15-13 EasyClip Xpress 12x15x13  EZMXP08-08-08 EasyClip Xpress 08x08x08  EZMXP10-10-10 EasyClip Xpress 10x10x10 EZMXP12-10-10 EasyClip Xpress 12x10x10 EZXP15-12-12 EasyClip Xpress 15x12x12  EZXP15-15-15 EasyClip Xpress 15x15x15  EZXP18-14-14 EasyClip Xpress 18x14x14 EZXP18-17-15 EasyClip Xpress 18x17x15 EZXP18-19-17 EasyClip Xpress 18x19x17 EZXP20-16-16 EasyClip Xpress 20x16x16 EZXP20-20-20 EasyClip Xpress 20x20x20 EZXP25-22-22 EasyClip Xpress 25x22x22

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0457-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0457-2021Class IIOngoingVoluntary: Firm initiated