cobas infinity central lab/cobas infinity core license-The cobas infinity IT solutions application is...
Roche Diagnostics Operations, Inc. began a recall of cobas infinity central lab/cobas infinity core license-The cobas infinity IT solutions application is intended to be used for the configuration and... on . FDA classified it as Class II and gives the reason as "Potential Incorrect Validation of Results Due to an Erroneous QC Status When Using Status Expiration Control Rule". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0461-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- cobas infinity central lab/cobas infinity core license-The cobas infinity IT solutions application is intended to be used for the configuration and connectivity management of instruments and software systems Catalog Number: 07154003001
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- Potential Incorrect Validation of Results Due to an Erroneous QC Status When Using Status Expiration Control Rule
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Roche issued Urgent Medical Device Correction TP-01100 is on 10/9/20 via UPS Ground (receipt signature required). Letter states reason for recall, health risk and action to take: Until cobas infinity Service Patches are available and implemented at your site, follow the containment measures outlined in the Applying Containment Measures section of this UMDC. " If you run the QC module in the cobas infinity application, contact the Roche Support Network Customer Support Center at 1-800-526-2272, and a representative will run a script to help you determine if your facility might be impacted. " Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. " Complete the attached faxback form (TP-01111) and fax or email it according to the instructions on the form. " File this Urgent Medical Device Correction (UMDC) for future reference. The cobas infinity Service Patches 3.01.12 (to correct affected versions 3.01.xx) and 3.02.05 (to correct affected versions 3.02.xx) are planned to be available in the fourth quarter of 2020. Once the Services Patches are available, a Roche Diagnostics representative will contact you to coordinate the corresponding installation.
- Quantity in commerce
- 40 licenses
- Distribution
- Nationwide
- Product codes and lots
- Software versions from cobas infinity version 3.00.00 onwards Serial Numbers: 82014 82229 82028 82227 82066 81967 81946 82121 82027 82194 82120 81528 82231 81958 82246 81983 82118 82165 81886 82082 82256 82257 81781 82331 82304 82087 82276 81465 81959 82235 82180 82181 82236 81945 82275 81662 82065 82253 81817 82000
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0461-2021 | Class II | Terminated | Voluntary: Firm initiated |