Z-0540-2021Class II

CORFLO NG/NI Pediatric Neonatal Feeding Tube w/ Anti-IV Connector

Halyard Health / initiated 2020-10-16 / Terminated

Avanos Medical, Inc. began a recall of CORFLO NG/NI Pediatric Neonatal Feeding Tube w/ Anti-IV Connector on . FDA classified it as Class II and gives the reason as "Detachment of the Male AIV Adaptor from the Y-Connector". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0540-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Halyard Health
Product
CORFLO NG/NI Pediatric Neonatal Feeding Tube w/ Anti-IV Connector
Product code
KNT Tubes, Gastrointestinal (And Accessories)
Reason for recall
Detachment of the Male AIV Adaptor from the Y-Connector.
FDA root cause
Under Investigation by firm
Action
On 10/16/2020 the firm sent a notification to its consignees with the following instructions: - CHECK all storage and usage locations to determine if any impacted product remains within your possession. Product Currently In-Use -DISCONTINUE using the impacted product (Note: If you decide not to replace the device, please consider taping over the adaptor to prevent any potential detachment and be sure to check on the tube and adaptor frequently throughout the day for potential detachment). -REPLACE the impacted product with a unit of a different lot. -IMMEDIATELY COMPLETE and RETURN the attached Recall Acknowledgement Form to Avanos . Via email to [email protected] -DESTROY the devices in your inventory following your facility's destruction procedures . Unused Inventory -SEGREGATE and QUARANTINE the devices according to your facility's procedures. -IMMEDIATELY COMPLETE and RETURN the attached Recall Acknowledgement Form to Avanos. Via email to fieldactioncare@avanos .com -DESTROY the devices in your inventory following your facility's destruction procedures.
Quantity in commerce
710 units
Distribution
Domestic distribution only: CA, CO, IL, LA, OH, VA.
Product codes and lots
Product Code:20-1225AIV, Lot Number: 0002983985

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0540-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0540-2021Class IITerminatedVoluntary: Firm initiated