Z-0540-2021Class II
CORFLO NG/NI Pediatric Neonatal Feeding Tube w/ Anti-IV Connector
Halyard Health / initiated 2020-10-16 / Terminated
Avanos Medical, Inc. began a recall of CORFLO NG/NI Pediatric Neonatal Feeding Tube w/ Anti-IV Connector on . FDA classified it as Class II and gives the reason as "Detachment of the Male AIV Adaptor from the Y-Connector". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0540-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Halyard Health
- Product
- CORFLO NG/NI Pediatric Neonatal Feeding Tube w/ Anti-IV Connector
- Product code
- KNT Tubes, Gastrointestinal (And Accessories)
- Reason for recall
- Detachment of the Male AIV Adaptor from the Y-Connector.
- FDA root cause
- Under Investigation by firm
- Action
- On 10/16/2020 the firm sent a notification to its consignees with the following instructions: - CHECK all storage and usage locations to determine if any impacted product remains within your possession. Product Currently In-Use -DISCONTINUE using the impacted product (Note: If you decide not to replace the device, please consider taping over the adaptor to prevent any potential detachment and be sure to check on the tube and adaptor frequently throughout the day for potential detachment). -REPLACE the impacted product with a unit of a different lot. -IMMEDIATELY COMPLETE and RETURN the attached Recall Acknowledgement Form to Avanos . Via email to [email protected] -DESTROY the devices in your inventory following your facility's destruction procedures . Unused Inventory -SEGREGATE and QUARANTINE the devices according to your facility's procedures. -IMMEDIATELY COMPLETE and RETURN the attached Recall Acknowledgement Form to Avanos. Via email to fieldactioncare@avanos .com -DESTROY the devices in your inventory following your facility's destruction procedures.
- Quantity in commerce
- 710 units
- Distribution
- Domestic distribution only: CA, CO, IL, LA, OH, VA.
- Product codes and lots
- Product Code:20-1225AIV, Lot Number: 0002983985
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0540-2021 | Class II | Terminated | Voluntary: Firm initiated |