Z-0399-2021Class I

Fixed Core Wire Guide, RPN TSCF-35-260-3, GPN G00517. The product is a curved wire guide with PTFE coating...

Cook Incorporated / initiated 2020-10-22 / Terminated / root cause: manufacturing process

Cook Inc. began a recall of Fixed Core Wire Guide, RPN TSCF-35-260-3, GPN G00517. The product is a curved wire guide with PTFE coating manufactured from stainless steel. To assist in... on . FDA classified it as Class I and gives the reason as "Affected product was dropped during manufacturing, which may result in a bent wire guide and/or the presence of loose foreign matter on the wire guide". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0399-2021
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Fixed Core Wire Guide, RPN TSCF-35-260-3, GPN G00517. The product is a curved wire guide with PTFE coating manufactured from stainless steel. To assist in inserting catheters, introducer sheaths, and other medical devices in vascular applications.
Product code
DQX Wire, Guide, Catheter
Reason for recall
Affected product was dropped during manufacturing, which may result in a bent wire guide and/or the presence of loose foreign matter on the wire guide.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On October 22, 2020, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were asked to take the following actions: 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email ([email protected]). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: [email protected].
Quantity in commerce
61 units
Distribution
US distribution to AZ, IA, IN, MA, and ND.
Product codes and lots
Lot 13199931 with UDI (01)00827002005174(17)250824(10)13199931

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0399-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0399-2021Class ITerminatedVoluntary: Firm initiated