Z-0463-2021Class II

MEDICREA LIGAPASS TENSION PULLEY HANDLE XS- intended to provide thoracic and/or lumbar posterior...

Medicrea International / initiated 2020-10-08 / Terminated / root cause: design

Medicrea International began a recall of MEDICREA LIGAPASS TENSION PULLEY HANDLE XS- intended to provide thoracic and/or lumbar posterior stabilization for degenerative pathologies Ref : A08200200 on . FDA classified it as Class II and gives the reason as "Tightening the LIGAPASS Band with this instrument, the tensioner's pliers can crush the band causing the band to break during surgery and lead to a risk of infection and re-intervention". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0463-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medicrea International
Product
MEDICREA LIGAPASS TENSION PULLEY HANDLE XS- intended to provide thoracic and/or lumbar posterior stabilization for degenerative pathologies Ref : A08200200
Product code
OWI Bone Fixation Cerclage, Sublaminar
Reason for recall
Tightening the LIGAPASS Band with this instrument, the tensioner's pliers can crush the band causing the band to break during surgery and lead to a risk of infection and re-intervention ,
Root cause tag
design
FDA root cause
Device Design
Action
Medicrea International issued Reportable voluntary Recall notification dated 10/5/20 to via e-mail to distributors and subsidiaries stating reason for recall, health risk and action to take: Instructions sent to US subsidiary : 1. Immediately check your internal inventory and quarantine all subject devices pending return to MEDICREA 2. Circulate this Field Safety Notice internally to all interested/affected parties. 3. Inform MEDICREA if any of the subject devices have been distributed to other organisations. a) Please provide contact details so that MEDICREA can inform the recipients appropriately. b) If you are a Distributor, note that you are responsible for notifying your affected customers. 4. Please inform MEDICREA of any adverse events concerning the use of the subject devices Please comply with any local laws or regulations concerning the notification of adverse events to your National Competent Authority. 5. Complete the attached customer response form. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also negate the need for us to send any further unnecessary communications on this matter. Therefore please complete even if you no longer have any of the subject devices in your physical inventory. Contact: Karine TROGNEUX MEDICREA INTERNATIONAL Regulatory Affairs Manager, Direct : +33(0)4.69.85.95.39 Standard : +33 (0)4.72.01.87.87. Fax - N +33 (0)04 72 01 87 88 Mail : [email protected]
Quantity in commerce
17 units US; 8 units OUS Expanded: 1 US and 3 OUS
Distribution
CA Foreign:France, Belgium
Product codes and lots
Lot Numbers: US: 17E0550-1, 17E1089R/2, 18E0278, OUS: 19B0962, 19B0595, 19B0601  Expanded Recall: 11/3/20 17E0550/2, 19B0602, 19B0963 and 19B0964

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0463-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0463-2021Class IITerminatedVoluntary: Firm initiated