ViewRay MRIdian Linac System: Model No. 20000, CE, Rated Supply Voltage - 380/480 VAC, 3 - Frequency Range...
Viewray, Inc. began a recall of ViewRay MRIdian Linac System: Model No. 20000, CE, Rated Supply Voltage - 380/480 VAC, 3 - Frequency Range (Hertz) - 50/60 Hz, Rated Input in Amperes - 210 A... on . FDA classified it as Class II and gives the reason as "There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0482-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Viewray, Inc.
- Product
- ViewRay MRIdian Linac System: Model No. 20000, CE, Rated Supply Voltage - 380/480 VAC, 3 - Frequency Range (Hertz) - 50/60 Hz, Rated Input in Amperes - 210 A, UDI. The MRIdian system and the MRIdian Linac system, with magnetic resonance imaging capabilities, is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 10/20/2020, the firm, ViewRay, Inc, emailed an "Urgent Device Correction" notification to all affected consignees. In addition to informing consignees about the correction, the firm asked consignees to take the following actions.: 1. For Customer using TPDS software ver. 5.2.5 or earlier with optional 3 Plane imaging: - Do: Maintain a distance of at least 5 cm (~2) between the receive coil and the inner surface of the bore wall when using 3 Plane cin treatment imaging (Figure 3 & 4). - Do Not: Select 3 Plane cin treatment imaging if a distance of at least 5 cm (~2) cannot be maintained between the receive coil and the inner surface of the bore wall during treatment. 2. For Customer Using TPDS software ver. 5.3 or later with optional 8 fps enabled: - Do Not: Use 8 fps cin treatment imaging until further guidance is provided by ViewRay. - Do: Use 4 fps cin treatment imaging when treating a patient. 3. If you have any questions related to this letter, please contact ViewRay Customer Support at [email protected] or call +1 855-286-8875.
- Quantity in commerce
- 35 units
- Distribution
- Worldwide Distribution: US (nationwide) to states of: CA, NY, FL, IL, MI, MO, OR, WI and OUS countries of: France, Germany, Italy, United Kingdom, Taiwan, Netherlands, Israel, and Denmark.
- Product codes and lots
- All Lots/Serial Numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0481-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0482-2021 | Class II | Terminated | Voluntary: Firm initiated |