Z-0471-2021Class III

VITROS Chemistry Products Performance Verifier II-For in vitro diagnostic use only.VITROS Chemistry Products...

Ortho-Clinical Diagnostics, Inc. / initiated 2020-10-16 / Open, Classified

Ortho Clinical Diagnostics Inc began a recall of VITROS Chemistry Products Performance Verifier II-For in vitro diagnostic use only.VITROS Chemistry Products Performance Verifier is an assayed control used to... on . FDA classified it as Class III and gives the reason as "Product instability; biased ALKP values obtained from VITROS Chemistry Products Performance Verifier II Lot Q7692 when stored unopened at refrigerated temperatures". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0471-2021
Classification
Class III
Status
Open, Classified
Initiated
Posted
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products Performance Verifier II-For in vitro diagnostic use only.VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated System - Product Code: 8231474
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
Product instability; biased ALKP values obtained from VITROS Chemistry Products Performance Verifier II Lot Q7692 when stored unopened at refrigerated temperatures
FDA root cause
Under Investigation by firm
Action
Ortho Clinical Diagnostics issued letter IMPORTANT PRODUCT CORRECTION NOTIFICATION (CL2020-243) on 16 October 2020 via FedEx/US Priority Mail or ORTHO Plus e-Communications for delivery on October 19, 2020, stating reason for recall, health risk and action to take: If their laboratory used VITROS Performance Verifier II Lot Q7692 to verify performance of VITROS ALKP Slides and stores an inventory of unopened vials in the refrigerator: Discard any remaining refrigerated inventory of unopened VITROS Performance Verifier II Lot Q7692. If their laboratory does not store inventory of VITROS Performance Verifier II Lot Q7692 in the refrigerator or does not use Lot Q7692 to verify performance of VITROS ALKP Slides: No Action Is Required. To verify performance of VITROS ALKP Slides, use frozen (never refrigerated) inventory of VITROS Performance Verifier II Lot Q7692 or switch to another lot of VITROS Performance Verifier II. Ortho will replace or credit your account as indicated on your Confirmation of Receipt form Retain this notification as verification of the revised instructions. OUS notified by email on 09/14/20 to global affiliates for distribution to distributors and customers.
Quantity in commerce
10788 total: US 4680; OUS 6108
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Product codes and lots
Lot Q7692 , EXP: 01-NOV-2021 UDI: 10758750004577

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0471-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0471-2021Class IIIOngoingVoluntary: Firm initiated