6PLUS1 Multilevel Saliva Calibrator Set Cortisol Cortisone in Saliva, Order No. 73039 -
Chromsystems Instruments & Chemicals GmbH began a recall of 6PLUS1 Multilevel Saliva Calibrator Set Cortisol Cortisone in Saliva, Order No. 73039 - Product Usage: The kit is used as a test for patients to monitor their... on . FDA classified it as Class II and gives the reason as "Calibrator lots may not meet stability criteria and could result in increased results of patient samples for cortisone". As of , the recall status on record is Completed, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0521-2021
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Chromsystems Instruments & Chemicals GmbH
- Product
- 6PLUS1 Multilevel Saliva Calibrator Set Cortisol Cortisone in Saliva, Order No. 73039 - Product Usage: The kit is used as a test for patients to monitor their cortisol and cortisone levels.
- Product code
- JIX Calibrator, Multi-Analyte Mixture
- Reason for recall
- Calibrator lots may not meet stability criteria and could result in increased results of patient samples for cortisone.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- The firm sent consignees a recall notification letter dated 10/28/2020. Consignees are asked to stop using products and to destroy remaining stock of affected product in accordance with the consignees laboratory procedures. The recalling firm will replace affected stock with new lots. Consignees are asked to further disseminate recall information to any other customers that the product has been shipped to. A response form is included with the recall notification letter and customers are asked to complete the form and return to the recalling firm.
- Quantity in commerce
- 148 package units
- Distribution
- US Nationwide distribution in the states of NC and NY.
- Product codes and lots
- Order No. 73039, Lot 3219 (expiration 08/01/2021) and Lot 1020 (Expiration 03/01/2022)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0521-2021 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-0522-2021 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-0523-2021 | Class II | Completed | Voluntary: Firm initiated |