GE Healthcare, LLC began a recall of MAC VU360, Electrocardiograph on . FDA classified it as Class II and gives the reason as "Incorrect patient identification and/or patient demographic errors". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0543-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- MAC VU360, Electrocardiograph
- Product code
- DPS Electrocardiograph
- Reason for recall
- Incorrect patient identification and/or patient demographic errors.
- FDA root cause
- Use error
- Action
- On 10/27/2020 the firm sent a letter to its consignees with the following instructions: You can continue to use your MAC VU360 system: At system setup: -Disable Auto-ECG in System Settings until the software has been upgraded. When performing ECGs: - Ensure the current patient lead wires are disconnected before selecting the Start New Patient workflow in the user interface. - Verify there are live scrolling waveforms on the screen before acquiring an ECG on the patient. - Verify the date and time of the ECG report corresponds to the patient for whom you are acquiring the ECG. - Always enter patient demographic data for each patient before acquiring an ECG GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the MAC VU360 software to be updated. Providing GE Healthcare with an email address in the attached customer response form, allows us to deliver this software update to you electronically and provide you with notifications of future software updates, as they become available. Note: After the MAC VU360 has been updated, discontinue usage of and destroy any media containing MAC VU360 V1.01 SP06 and any prior versions of MAC VU360 software. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Quantity in commerce
- 5,284 units
- Distribution
- Domestic Distribution: AL,AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI,MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. International Distribution: AUSTRALIA, AUSTRIA, BELGIUM, Canada, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, Germany, Hong Kong, IRELAND, ISRAEL, ITALY, Korea, KUWAIT, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, QATAR, SINGAPORE, SPAIN, SWEDEN, SWITZERLAND, THAILAND, UNITED ARAB EMIRATES, United Kingdom.
- Product codes and lots
- Model: 2030360 001
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0543-2021 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
