Sentinel V10.x & V11.x, Models: 98200 & 98201 -
Spacelabs Healthcare, Ltd. began a recall of Sentinel V10.x & V11.x, Models: 98200 & 98201 - Product Usage: designed using the latest programming technology and is easy to install and to use on . FDA classified it as Class II and gives the reason as "During system upgrade incorrect comments may be added to existing tests. In such instance: Resting ECG and ABP tests which did not contain comments at the time of the upgrade, may have unrelated comments added. Existing...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0563-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Sentinel V10.x & V11.x, Models: 98200 & 98201 - Product Usage: designed using the latest programming technology and is easy to install and to use.
- Product code
- DQK Computer, Diagnostic, Programmable
- Reason for recall
- During system upgrade incorrect comments may be added to existing tests. In such instance: Resting ECG and ABP tests which did not contain comments at the time of the upgrade, may have unrelated comments added. Existing comments will not be overwritten or removed. Rhythm ECG Tests may have comments overwritten with unrelated comments. Any tests created after the upgrade are not affected.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 11/06/20, Urgent Medical Device Correction notices were mailed to customers. Customers were asked to do the following: Any comments associated with ABP, Resting, or Rhythm tests that existed in your system prior to the date of your upgrade should be reviewed for accuracy before being used to inform any clinical decision. Weigh the benefits versus the risks when deciding whether to use the comments associated with ABP, Resting, or Rhythm tests which existed in your system prior to the date of your upgrade of Sentinel. We recommend that you brief your staff regarding the issue listed above. Complete and return the customer reply form or use the following link to acknowledge receipt of this letter: https://www.spacelabshealthcare.com/support/recall-reply-form/ For additional information or technical assistance, please contact: Global Technical Support at, +44 (0) 1992 507740, or [email protected]
- Quantity in commerce
- 600
- Distribution
- Worldwide distribution - U.S. Nationwide distribution including in the states of PR, MN, NE, FL, IL, CA, NY, ND, AR, WI, SD, TN, MO, MA, TX, MT, OH, MI, IA, NJ, CT, NC, UT, WV, GA, CO, PA, AL, DC, NV, DE, WA, OR, IN, SC, HI, MS, KS, ID, MD, LA, NH, VA, ME, AK, VT, OK, AZ, KY, NM, RI and the countries of AFGHANISTAN, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BELGIUM, BOLIVIA, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, ICELAND, INDIA, IRAN, IRELAND, ISRAEL, ITALY, KENYA, KOREA, REPUBLIC OF, KUWAIT, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, OMAN, PAKISTAN, PALESTINE, PANAMA, PARAGUAY, PERU, POLAND, PORTUGAL, QATAR, Republic of Macedonia, SAUDI ARABIA, Serbia, SINGAPORE, Slovakia, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VENEZUELA.
- Product codes and lots
- Systems that have been upgraded from Version 9 or below to Version 10 or above.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0563-2021 | Class II | Terminated | Voluntary: Firm initiated |