Z-0757-2021Class II
Contour VL Variable Length Ureteral Stent, UPN M0061801550, M0061801560, M0061801570
Boston Scientific Corporation / initiated 2020-11-16 / Open, Classified / root cause: manufacturing process
Boston Scientific Corporation began a recall of Contour VL Variable Length Ureteral Stent, UPN M0061801550, M0061801560, M0061801570 on . FDA classified it as Class II and gives the reason as "Trace amounts of metallic debris were found on the stent related to the manufacturing process. The debris was confirmed to be particles of nickel". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0757-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Boston Scientific Corporation
- Product
- Contour VL Variable Length Ureteral Stent, UPN M0061801550, M0061801560, M0061801570
- Product code
- FAD Stent, Ureteral
- Reason for recall
- Trace amounts of metallic debris were found on the stent related to the manufacturing process. The debris was confirmed to be particles of nickel.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- On November 16, 2020, the firm distributed Urgent Medical Device Removal letters to affected customers. Customers were informed of the product issue and asked to take the following actions: - Verify whether you have any product from the affected lots/batches within your inventory - Immediately cease use or distribution of any remaining unused product from the affected lots/batches - Segregate affected material and return it to BSC in accordance with removal instructions included in the letter. - Distributors should forward the notification to their customers Your local Sales Representative can answer any questions that you may have regarding this notification.
- Quantity in commerce
- 458 units
- Distribution
- Distribution US nationwide and Japan.
- Product codes and lots
- GTIN 08714729058410: Lots 25578577, 25598812, 25609220, 25614542, 25614545, 25619687, 25641771, 25713852, 25753933, 25790349. GTIN 08714729067986: Lots 25640715, 25670518, 25686790, 25686791, 25686792, 25741763, 25741770, 25753937, 25790351, 25795704, 25802759, 25809481, 25822104. GTIN 08714729068020: Lots 25578740, 25583824.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0757-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0758-2021 | Class II | Ongoing | Voluntary: Firm initiated |