Z-0721-2021Class II

ADVIA Gamma-Glutamyl Transferase (GGT) Reagent -

Siemens Healthcare Diagnostics Inc. / initiated 2020-11-23 / Open, Classified / root cause: design

Siemens Healthcare Diagnostics, Inc. began a recall of ADVIA Gamma-Glutamyl Transferase (GGT) Reagent - Product Usage: For in vitro diagnostic use in the quantitative determination of gamma-glutamyl transferase in... on . FDA classified it as Class II and gives the reason as "Gamma-Glutamyl Transferase (GGT) Reagent Low End Repeatability-Imprecision. Reagent may demonstrate a coefficient of variation (%CV) outside of the performance data for repeatability (within-run) listed in the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0721-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ADVIA Gamma-Glutamyl Transferase (GGT) Reagent - Product Usage: For in vitro diagnostic use in the quantitative determination of gamma-glutamyl transferase in human serum and plasma. Siemens Material Number (SMN): ADVIA Chemistry: 10309495 (20 mL) 10316298 (70 mL)
Product code
JQA Electrophoretic, Gamma-Glutamyl Transpeptidase Isoenzymes
Reason for recall
Gamma-Glutamyl Transferase (GGT) Reagent Low End Repeatability-Imprecision. Reagent may demonstrate a coefficient of variation (%CV) outside of the performance data for repeatability (within-run) listed in the Instructions For Use (IFU) on some analyzers.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens Healthcare Diagnostics Urgent Field Safety Notice letter beginning November 23, 2020 via Fed Ex and outside the US the distribution is determined at the country level. Letter states reason for recall , health risk and action to take: For the product listed above, please perform the following steps: 1. Review and apply the preliminary precision performance characteristics of d8% CV at a GGT concentration of approximately 27 U/L 42 U/L provided. 2. Please review this letter with your Medical Director. 3. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. 4. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Customer Care Center or your local Siemens technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Contact your Siemens Customer Care Center or your local Siemens technical support representative. The precision section of the IFUs will be updated when the investigation is complete. The updated IFUs will be uploaded into Document Library where all registered users who opt in to receive alerts will be notified of the updated IFU.
Quantity in commerce
397 units US; 1713 units OUS
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Australia, Austria, Bahamas, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, India, Iran, Ireland, Israel, Italy, Japan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., United Kingdom, Uruguay, Vatican City, Vietnam. ¿¿¿¿ ¿¿¿¿¿
Product codes and lots
All lot codes. UDI: 10309495 (20 mL) = 00630414508344 10316298 (70 mL) = 00630414209753

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0721-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0721-2021Class IIOngoingVoluntary: Firm initiated
Z-0722-2021Class IIOngoingVoluntary: Firm initiated