Percept PC Implantable Neurostimulator (INS) -
Medtronic Neuromodulation began a recall of Percept PC Implantable Neurostimulator (INS) - Product Usage: The Percept PC is a neurostimulation device for deep brain stimulation therapy. It is indicated... on . FDA classified it as Class II and gives the reason as "A software anomaly in the A620 Patient Programmer application was identified that results in failure to connect with the Percept PC device". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0612-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Neuromodulation
- Product
- Percept PC Implantable Neurostimulator (INS) - Product Usage: The Percept PC is a neurostimulation device for deep brain stimulation therapy. It is indicated for DBS treatment for Parkinson s disease, essential tremor, dystonia, obsessive-compulsive disorder, and epilepsy.
- Product code
- MHY Stimulator, Electrical, Implanted, For Parkinsonian Tremor
- Reason for recall
- A software anomaly in the A620 Patient Programmer application was identified that results in failure to connect with the Percept PC device.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Beginning November 19, 2020 accounts impacted by this issue were verbally notified to quarantine affected product. Beginning November 25, 2020 a recall notification letter and retrieval instructions titled "URGENT: MEDICAL DEVICE RECALL A620 Patient Programmer issue with specific Percept" PC Neurostimulators" was hand delivered to impacted accounts by Medtronic Representatives.
- Quantity in commerce
- 12 devices
- Distribution
- US Nationwide distribution including in the states of Arizona, California, New Hampshire, Pennsylvania, Tennessee, Texas, Washington.
- Product codes and lots
- Model Number: B35200; GTIN: 00763000253363; Serial Numbers: NPI706941H, NPI707024H, NPI707000H, NPI706992H, NPI706963H, NPI706962H, NPI706987H, NPI706945H, NPI706952H, NPI706954H, NPI706955H, NPI706957H
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0612-2021 | Class II | Terminated | Voluntary: Firm initiated |