Emission Computed Tomography System Image Process System -
Philips North America, LLC began a recall of Emission Computed Tomography System Image Process System - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of... on . FDA classified it as Class II and gives the reason as "A software defect that has the potential to result in image misdiagnosis and incorrect treatment of a patient". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0852-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Emission Computed Tomography System Image Process System - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- A software defect that has the potential to result in image misdiagnosis and incorrect treatment of a patient.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 11/23/2020, Philips issued Urgent Device Correction notices to customers via press release, letter, telefax, telephone, e-mail, visit, etc. Due to a software defect Philips is notifying all affected users of the JETStream Workspace Thyroid Analysis application recommending that customers calculate thyroid uptake utilizing the method provided in the notices and not using the Transfer Coefficient parameter.
- Quantity in commerce
- 2714
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Canada, Mexico, Algeria, Argentina, Australia, Bahrain, Austria, Belgium, Brazil, China, Columbia, Cuba, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guatemala, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Kuwait, Latvia, Libya, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Kingdom, Venezuela.
- Product codes and lots
- Model# 882310,882311,882313 All Lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0852-2021 | Class II | Terminated | Voluntary: Firm initiated |