Z-0852-2021Class II

Emission Computed Tomography System Image Process System -

Philips North America LLC / initiated 2020-11-19 / Terminated / root cause: design

Philips North America, LLC began a recall of Emission Computed Tomography System Image Process System - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of... on . FDA classified it as Class II and gives the reason as "A software defect that has the potential to result in image misdiagnosis and incorrect treatment of a patient". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0852-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Emission Computed Tomography System Image Process System - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
A software defect that has the potential to result in image misdiagnosis and incorrect treatment of a patient.
Root cause tag
design
FDA root cause
Device Design
Action
On 11/23/2020, Philips issued Urgent Device Correction notices to customers via press release, letter, telefax, telephone, e-mail, visit, etc. Due to a software defect Philips is notifying all affected users of the JETStream Workspace Thyroid Analysis application recommending that customers calculate thyroid uptake utilizing the method provided in the notices and not using the Transfer Coefficient parameter.
Quantity in commerce
2714
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Mexico, Algeria, Argentina, Australia, Bahrain, Austria, Belgium, Brazil, China, Columbia, Cuba, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guatemala, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Kuwait, Latvia, Libya, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Kingdom, Venezuela.
Product codes and lots
Model# 882310,882311,882313 All Lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0852-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0852-2021Class IITerminatedVoluntary: Firm initiated