Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0723-2021Class II

Vanguard 360 Revision System Posterior Augment With Bolt -

Zimmer GmbH / initiated 2020-12-01 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of Vanguard 360 Revision System Posterior Augment With Bolt - Product Usage: intended for use in knee arthroplasty Item Number: 185422 on . FDA classified it as Class II and gives the reason as "Incorrect bolt -tibial bolts in package instead of femoral bolts may delay procedure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0723-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Vanguard 360 Revision System Posterior Augment With Bolt - Product Usage: intended for use in knee arthroplasty Item Number: 185422
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Incorrect bolt -tibial bolts in package instead of femoral bolts may delay procedure
Root cause tag
manufacturing process
FDA root cause
Process control
Action
.Zimmer Biomet issued Urgent Medical Device Recall letter issued on 12/1/20 to distributors with direct sales and/or fielded inventory via email and traceable letter sent via FedEx. Hospitals with direct sales or current inventory will be notified via FedEx. Letter states reason for recall , health risk and action to take: Distributors esponsibilities include: locating and removing the product in their territory, as well as identifying hospitals who have previously used the product. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected].
Quantity in commerce
21 units
Distribution
Worldwide distribution - US Nationwide distribution in the state of FL and the countries of Korea, Japan, Netherlands.
Product codes and lots
Lot Number: 098780  UDI Number: (01) 00880304482739 (17) 290501 (10) 098780

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0723-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0723-2021Class IITerminatedVoluntary: Firm initiated