Z-0805-2021Class II

Uroskop Omnia, Model no. 10094910 -

Siemens Medical Solutions USA, Inc. / initiated 2020-09-09 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Uroskop Omnia, Model no. 10094910 - Product Usage: a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional x-ray... on . FDA classified it as Class II and gives the reason as "After the March 2019 corrective action on the monitor support arm, there is a residual risk of the support arm becoming dislodged in cases when the monitor support arm is used intensely over a long period of time". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0805-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Uroskop Omnia, Model no. 10094910 - Product Usage: a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional x-ray diagnosis, endourology and minimal invasive urology/surgery).
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
After the March 2019 corrective action on the monitor support arm, there is a residual risk of the support arm becoming dislodged in cases when the monitor support arm is used intensely over a long period of time.
Root cause tag
design
FDA root cause
Device Design
Action
On September 9, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that the firm has released an additional safety measure to avoid the potential issue. A safety gear will be installed into the monitor support arm out of an abundance of caution. The corrective action is available and is being provided to affected users free of charge. The firm will contact customers to arrange a date to perform the corrective action. Please feel free to contact the firm to schedule an earlier appointment at 1-800-888-7436.
Quantity in commerce
6
Distribution
Worldwide distribution.
Product codes and lots
Serial Nos. 3417 3414 3415 3418 3428 3416