Z-0944-2021Class II

Philips EPIQ Diagnostic Ultrasound System; Model Numbers: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 7G, EPIQ 7C...

Philips North America LLC / initiated 2020-12-16 / Terminated / root cause: manufacturing process

Philips Ultrasound Inc began a recall of Philips EPIQ Diagnostic Ultrasound System; Model Numbers: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 7G, EPIQ 7C, EPIQ CVx & EPIQ CVxi. MOD EPIQ Elite DIAGNOSTIC... on . FDA classified it as Class II and gives the reason as "The manufacturer discovered an issue associated with the EPIQ Image Boost with xPlane and Color Flow or Doppler while using the X8-2t TEE Transducer. If Image Boost is enabled the color box shows flow visualization not...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0944-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips EPIQ Diagnostic Ultrasound System; Model Numbers: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 7G, EPIQ 7C, EPIQ CVx & EPIQ CVxi. MOD EPIQ Elite DIAGNOSTIC ULTRASOUND SYSTEM; Rx Only 100-240V~50/60 Hz, 600 VA MOD EPIQ 5 DIAGNOSTIC ULTRASOUND SYSTEM; GMDN 40761, Rx Only, 100-240V~50/60 Hz, 600VA MOD EPIQ 7 DIAGNOSTIC ULTRASOUND SYSTEM; GMDN 40761, Rx Only, 100-240V~50/60 Hz, 600VA MOD EPIQ DIAGNOSTIC ULTRASOUND SYSTEM; Rx Only,100-240V~50/60 Hz, 600VA
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
The manufacturer discovered an issue associated with the EPIQ Image Boost with xPlane and Color Flow or Doppler while using the X8-2t TEE Transducer. If Image Boost is enabled the color box shows flow visualization not accurately represent the fluid flow. Similarly, if Image Boost is enabled, both CW and PW Doppler traces will not accurately represent the fluid flow.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
On 12/16/2020, the manufacturer sent an Urgent Medical Device Correction letter (MDC) with instructions on how to turn off the Image Boost feature. Philips will resolve the issue by providing a software update, at no cost, to correct all affected EPIQ Ultrasound Systems (the update will disable the Image Boost feature when connected to an X8-2t TEE Transducer). Method of Notification - Letter (Urgent Medical Device Correction Letter in the US, Field Safety Notice in rest of the world, Distributor Letter in rest of the world). For US customers, the letters will be sent via certify mail with electronic return receipt. Philips Markets Organizations are responsible for distributing the letters outside of the US.
Quantity in commerce
607 units
Distribution
US - AL, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE,NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, WA, WI and WV OUS- Austria, Australia, Belgium, Canada, Chezh Republic, Croatia, Denmark, Egypt, Finland, France, Hong Kong, Indonesia, India, Ireland, Italy, Japan, Latvia, Lithuania, Malta, Mexico, Netherlands, Norway, Poland, Romania, Russia, Senegal, Slovenia, Sri Lanka, South Africa, Switzerland, Thailand, Turkey, UAE and United Kingdom.
Product codes and lots
All EPIQ Systems with software version 7.0 Model Numbers: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 7G, EPIQ 7C, EPIQ CVx & EPIQ CVxi

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0944-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0944-2021Class IITerminatedVoluntary: Firm initiated