Z-0939-2021Class II

FLOW COUPLER Monitor Flow Coupler devices are sold in 5 sizes (2.0 mm, 2.5 mm, 3.0 mm, 3.5 mm, 4.0 mm)...

Baxter Healthcare Corporation / initiated 2020-12-28 / Open, Classified / root cause: software

Baxter Healthcare Corporation began a recall of FLOW COUPLER Monitor Flow Coupler devices are sold in 5 sizes (2.0 mm, 2.5 mm, 3.0 mm, 3.5 mm, 4.0 mm). Implantable device for vascular use on . FDA classified it as Class II and gives the reason as "Potential for the battery to lose its ability to be recharged". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0939-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Baxter Healthcare Corporation
Product
FLOW COUPLER Monitor Flow Coupler devices are sold in 5 sizes (2.0 mm, 2.5 mm, 3.0 mm, 3.5 mm, 4.0 mm). Implantable device for vascular use.
Product code
MVR Device, Anastomotic, Microvascular
Reason for recall
Potential for the battery to lose its ability to be recharged.
Root cause tag
software
FDA root cause
Software design
Action
Written notification letter titled, "Urgent Medical Device Recall" was sent to all consignees on 12/28/2020 via first class mail. The letter instructs the consignee to locate all affected devices at their facility; contact Synovis MCA to arrange for product return for the firmware upgrade, power button update, and battery replacement, if applicable; complete the enclosed Baxter customer reply form and return it to Baxter by e-mailing it to [email protected], even if the consignees does not have any inventory; and if the consignee distributes this product to other facilities or departments within their institution, to please forward a copy of this communication to them.
Quantity in commerce
109 units
Distribution
Distributed nationwide to AZ, FL, OH, NY, GA, VA, MA, WI, MO, MT, WV, IL, NC, NJ, AR, CA, CO, CT, IA, ID, IN, KS, LA, MD, MI, MN, MS, NE, NH, NM, NV, OK, OR, PA, RI, WY, WA, VT, TX, TN, SC, SD
Product codes and lots
UDI: 00844735006148; Product Code: 5156-00000-011; Serial Numbers: 135306000279, 135306000280, 135306000281, 135306000282, 135306000283, 135306000284, 135306000286, 135306000287, 135306000288, 135306000289, 135306000290, 135306000291, 135306000292, 135306000293, 135306000294, 135306000295, 135306000298, 135306000299, 135306000300, 135306000301, 135306000303, 135306000304, 135306000305, 135591000306, 135591000307, 136959000309, 136959000310, 136959000311, 136959000312, 136959000313, 136959000314, 136959000315, 136959000316, 136959000317, 136959000318, 136959000319, 136959000320, 136959000321, 136959000322, 136959000323, 136959000324, 136959000325, 136959000326, 136959000327, 136959000328, 136959000329, 136959000330, 136959000331, 136959000332, 136959000333, 136959000334, 136959000335, 136959000336, 136959000337, 136959000338, 136959000339, 136959000340, 136959000341, 136959000342, 136959000344, 136959000345, 136959000346, 136959000347, 136959000348, 136959000349, 136959000350, 136959000351, 136959000352, 136959000353, 136959000354, 136959000355, 136959000357, 136959000358, 136959000359, 136959000360, 136959000361, 136959000362, 136959000364, 136959000365, 136959000366, 137645000392, 137645000393, 137645000394, 137645000395, 137645000396, 137645000397, 137645000399, 137645000400, 137645000401, 139608000552, 139608000555, 139608000571, 139609000583, 139609000589, 139609000596, 139609000597, 139609000598, 139609000599, 139609000600, 139609000601, 139609000602, 139609000603, 139609000604, 139609000605, 139609000606, 139609000607, 139609000608, 139609000609, 139609000610

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0939-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0939-2021Class IIOngoingVoluntary: Firm initiated