Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0854-2021Class II

GE NM/CT 860 system Model Number H3908AD -

GE Healthcare / initiated 2020-12-10 / Open, Classified / root cause: manufacturing process

GE Healthcare, LLC began a recall of GE NM/CT 860 system Model Number H3908AD - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake... on . FDA classified it as Class II and gives the reason as "GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0854-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
GE NM/CT 860 system Model Number H3908AD - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm initiated their action by letter to the consignees on 12/10/2020. The letter explained the issue and stated that GE would inspect, and if required, correct the affected units. Questions and concerns may be directed to GE Healthcare Service at 1-800-437-1171 or the consignee's local Service Representative.
Quantity in commerce
12 units
Distribution
Worldwide
Product codes and lots
Serial numbers: 860Y62016, 860Y62019, 860Y62017, 860Z62032, 860Y62018, 860Y62015, 860Y62012, 860Y62020, 860Y62021, 860Y62013,  860Z62033, 860Z62031

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0854-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0853-2021Class IITerminatedVoluntary: Firm initiated
Z-0854-2021Class IITerminatedVoluntary: Firm initiated
Z-0855-2021Class IITerminatedVoluntary: Firm initiated
Z-0856-2021Class IITerminatedVoluntary: Firm initiated
Z-0857-2021Class IITerminatedVoluntary: Firm initiated
Z-0858-2021Class IITerminatedVoluntary: Firm initiated