Z-0857-2021Class II

Discovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Model 3100XB)...

GE Healthcare / initiated 2020-12-10 / Open, Classified / root cause: manufacturing process

GE Healthcare, LLC began a recall of Discovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Model 3100XB), Discovery 670 ES (Model H3100XF) - Product Usage... on . FDA classified it as Class II and gives the reason as "GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0857-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
Discovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Model 3100XB), Discovery 670 ES (Model H3100XF) - Product Usage: intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm initiated their action by letter to the consignees on 12/10/2020. The letter explained the issue and stated that GE would inspect, and if required, correct the affected units. Questions and concerns may be directed to GE Healthcare Service at 1-800-437-1171 or the consignee's local Service Representative.
Quantity in commerce
52 untits
Distribution
Worldwide
Product codes and lots
Discovery 670 DR (Model H3100BT), Serial Numbers: DRGY56033, DRGY56037, DRGY56035, DRGY56030, DRGZ56043, DRGZ56042, DRGY56032, DRGY56031, DRGY56034, DRGY56036    NM/CT 870 DR (Model 3100Q), Serial Numbers: 870Y64066, and (Model H3100AS), Serial Numbers: 870Y64077, 870Y64064, 870Y64081, 870Y64079, 870Y64067, 870Y64074, 870Z64117, 870Y64078, 870Y64080, 870Y64069, 870Y64085, 870Z64121, 870Z64119, 870Z64122, 870Y64056, 870Y64063, 870Z64114, 870Z64115, 870Y64068, 870Y64082, 870Z64116, 870Z64120, 870Y64065, 870Y64084, 870Y64061, 870Y64083, 870Z64113, 870Y64070, 870Z64123, 870Y64058, 870Z64118, 870Y64057   Discovery NM670 Pro (Model 3100XB), Serial Numbers: PRGY52027,  PRGY52031, PRGZ52047, PRGZ52046, PRGY52028, PRGZ52048, PRGY52029, PRGY52030    Discovery 670 ES (Model H3100XF), Serial Numbers: ESGZ53005

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0857-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0853-2021Class IITerminatedVoluntary: Firm initiated
Z-0854-2021Class IITerminatedVoluntary: Firm initiated
Z-0855-2021Class IITerminatedVoluntary: Firm initiated
Z-0856-2021Class IITerminatedVoluntary: Firm initiated
Z-0857-2021Class IITerminatedVoluntary: Firm initiated
Z-0858-2021Class IITerminatedVoluntary: Firm initiated