Optima NM/CT 640, Model number 3101RL -
GE Healthcare, LLC began a recall of Optima NM/CT 640, Model number 3101RL - Product Usage: intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in... on . FDA classified it as Class II and gives the reason as "GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0858-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- Optima NM/CT 640, Model number 3101RL - Product Usage: intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm initiated their action by letter to the consignees on 12/10/2020. The letter explained the issue and stated that GE would inspect, and if required, correct the affected units. Questions and concerns may be directed to GE Healthcare Service at 1-800-437-1171 or the consignee's local Service Representative.
- Quantity in commerce
- 7 units
- Distribution
- Worldwide
- Product codes and lots
- Serial Numbers: 64GY51017, 64GY51020, 64GY51022, 64GZ51028, 64GZ51030, 64GZ51029, 64GY51021
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0853-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0854-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0855-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0856-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0857-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0858-2021 | Class II | Terminated | Voluntary: Firm initiated |