Artis zee/ zeego & Artis Q/ Q.zen Artis Pheno
Siemens Medical Solutions USA, Inc began a recall of Artis zee/ zeego & Artis Q/ Q.zen Artis Pheno on . FDA classified it as Class II and gives the reason as "When the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro "Fluoro Override", the measured Air Kerma Rate can exceed the regulatory limits". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0656-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Artis zee/ zeego & Artis Q/ Q.zen Artis Pheno
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- When the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro "Fluoro Override", the measured Air Kerma Rate can exceed the regulatory limits.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice to all affected customers via AX067/20/P on November 30, 2020. All modifications and repairs are provided to affected customers free of charge.
- Quantity in commerce
- 5602
- Distribution
- Worldwide-US Nationwide
- Product codes and lots
- Model number: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Pheno- 10849000 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis Q.zen floor- 10848353 Artis Q.zen ceiling- 10848354 Artis Q.zen biplane- 10848355
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0656-2021 | Class II | Ongoing | FDA Mandated |