Z-0656-2021Class II

Artis zee/ zeego & Artis Q/ Q.zen Artis Pheno

Siemens Medical Solutions USA, Inc. / initiated 2020-11-30 / Open, Classified / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Artis zee/ zeego & Artis Q/ Q.zen Artis Pheno on . FDA classified it as Class II and gives the reason as "When the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro "Fluoro Override", the measured Air Kerma Rate can exceed the regulatory limits". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0656-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artis zee/ zeego & Artis Q/ Q.zen Artis Pheno
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
When the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro "Fluoro Override", the measured Air Kerma Rate can exceed the regulatory limits.
Root cause tag
software
FDA root cause
Software design
Action
Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice to all affected customers via AX067/20/P on November 30, 2020. All modifications and repairs are provided to affected customers free of charge.
Quantity in commerce
5602
Distribution
Worldwide-US Nationwide
Product codes and lots
Model number: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Pheno- 10849000 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis Q.zen floor- 10848353 Artis Q.zen ceiling- 10848354 Artis Q.zen biplane- 10848355

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0656-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0656-2021Class IIOngoingFDA Mandated