Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0933-2021Class II

VERSABOND AB 40 GRAMS FORMULATION 2, antibiotic infused bone cement, Product Number: 71271440 -

Smith & Nephew, Inc. / initiated 2020-12-18 / Terminated

Smith & Nephew, Inc. began a recall of VERSABOND AB 40 GRAMS FORMULATION 2, antibiotic infused bone cement, Product Number: 71271440 - Product Usage: VERSABOND is a medium viscosity bone cement... on . FDA classified it as Class II and gives the reason as "Potential heat and humidity exposure. The affected lot contained an atypical sandy, flaky or grainy consistency when mixing". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0933-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
VERSABOND AB 40 GRAMS FORMULATION 2, antibiotic infused bone cement, Product Number: 71271440 - Product Usage: VERSABOND is a medium viscosity bone cement which can be used for hip, knee, and shoulder implant fixation. VERSABOND AB includes the antibiotic- gentamycin.
Product code
LOD Bone Cement
Reason for recall
Potential heat and humidity exposure. The affected lot contained an atypical sandy, flaky or grainy consistency when mixing.
FDA root cause
Storage
Action
A Recall notification letter titled, "Urgent Medical Device Recall Notice" was sent to consignees on 12/18/2020 via email and FedEx. The letter instructs the consignee to follow the instructions on the attached response form. The response form provides the following instructions: "1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customers account number and name in the table below. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to [email protected]. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact [email protected]."
Quantity in commerce
5190 devices
Distribution
Worldwide distribution - US Nationwide distribution in the state of OH and the countries of Dubai, Canada, Korea, Chile, Puerto Rico, Panama, and Costa Rica.
Product codes and lots
Product Number: 71271440; Batch Number: 18LC07540

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0933-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0933-2021Class IITerminatedVoluntary: Firm initiated