Z-0727-2021Class II

ProxiDiagnost N90 Image-intensified fluoroscopic x-ray system

Philips North America LLC / initiated 2020-12-08 / Terminated

PHILIPS HOME HEALTHCARE SOLUTION began a recall of ProxiDiagnost N90 Image-intensified fluoroscopic x-ray system on . FDA classified it as Class II and gives the reason as "After the user selects the Lock-in command, the kV and mA values are not locked immediately and can adjust if the collimator shutter position is changed directly after the Lock-in command". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0727-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
ProxiDiagnost N90 Image-intensified fluoroscopic x-ray system
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
After the user selects the Lock-in command, the kV and mA values are not locked immediately and can adjust if the collimator shutter position is changed directly after the Lock-in command.
FDA root cause
Radiation Control for Health and Safety Act
Quantity in commerce
91
Distribution
Nationwide
Product codes and lots
Pending

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0727-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0727-2021Class IITerminatedFDA Mandated