SenTec Digital Monitor, Model SDM, is a portable stand-alone patient monitor indicated for continuous...
SenTec AG began a recall of SenTec Digital Monitor, Model SDM, is a portable stand-alone patient monitor indicated for continuous, noninvasive patient monitoring of carbon dioxide partial... on . FDA classified it as Class II and gives the reason as "Device requires the use of an Isolation Transformer in combination with the SDM in home care settings to comply with applicable safety standards". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1134-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Sentec AG
- Product
- SenTec Digital Monitor, Model SDM, is a portable stand-alone patient monitor indicated for continuous, noninvasive patient monitoring of carbon dioxide partial pressure (PCo2), oxygen partial pressure (POs), functional oxygen saturation (PsO2), and pulse rate (PR), using either a single digital sensor (V-Sign Sensor 2 or OxiVenT sensor).
- Product code
- LKD Monitor, Carbon-Dioxide, Cutaneous
- Reason for recall
- Device requires the use of an Isolation Transformer in combination with the SDM in home care settings to comply with applicable safety standards.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm initiated their correction by letter to the direct accounts on 12/10/2020. The letter explains the problem and provides additional instruction manuals and a sticker to address the use of the product in the home care environment.
- Quantity in commerce
- 15 devices
- Distribution
- IL, PA, FL, NJ, GA, MA
- Product codes and lots
- Serial Numbers: 303170, 311263, 311166, 311408, 303426* *serial number information current as of today's date