Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1102-2021Class II

Baxter ONE-LINK Non-DEHP Y-Type Microbore Catheter Extension Set, Bonded Needle-free IV Connector, Neutral...

Baxter Healthcare Corporation / initiated 2020-12-21 / Terminated / root cause: manufacturing process

Baxter Healthcare Corporation began a recall of Baxter ONE-LINK Non-DEHP Y-Type Microbore Catheter Extension Set, Bonded Needle-free IV Connector, Neutral Fluid Displacement, 5.7 " (14 cm), REF 7N8370K -... on . FDA classified it as Class II and gives the reason as "There were customer reports of separation between the tubing and Male Luer Lock Adaptor". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1102-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Baxter ONE-LINK Non-DEHP Y-Type Microbore Catheter Extension Set, Bonded Needle-free IV Connector, Neutral Fluid Displacement, 5.7 " (14 cm), REF 7N8370K - Product Usage: intended for single patient use with a vascular access device for the administration of drugs and solutions without needles, thus eliminating the potential for needle-stick injuries during use.
Product code
FPA Set, Administration, Intravascular
Reason for recall
There were customer reports of separation between the tubing and Male Luer Lock Adaptor.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
The firm initiated the recall by email on 12/21/2020. The letter requested the following: 1. Locate and remove the affected product from consignee's facility. The product code and lot number can be found on the individual product or shipping carton. 2. Contact Baxter Healthcare Center for Service to arrange for return and credit. 3. Confirm the receipt of the letter by completing the enclosed Baxter customer reply form and return it to Baxter via email
Quantity in commerce
66,800 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of SC, OH, MI, CA, TX, and the countries of UK, Ireland.
Product codes and lots
UDI: 00085412613505, Lot Numbers: DR18E01023, exp. 5/2/2023; DR18E07053, exp. 5/8/2023; DR19G08023, exp. 7/8/2024; DR19H07015, exp. 8/7/2024; DR19L09030, exp. 12/10/2024; DR20B04020, exp. 2/5/2025

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1102-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1102-2021Class IITerminatedVoluntary: Firm initiated