Z-1101-2021Class II

GE Healthcare Precision 600FP, Classical R&F (radiographic and fluoroscopic) System -

GE Healthcare / initiated 2020-12-10 / Terminated

GE Healthcare, LLC began a recall of GE Healthcare Precision 600FP, Classical R&F (radiographic and fluoroscopic) System - Product Usage: Intended for use by a qualified/trained doctor or... on . FDA classified it as Class II and gives the reason as "GE Healthcare has become aware of a potential safety issue associated with the improper installation of the Precision 600FP monitor suspension. A partial detachment of the monitor suspension could result in potential...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1101-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Healthcare Precision 600FP, Classical R&F (radiographic and fluoroscopic) System - Product Usage: Intended for use by a qualified/trained doctor or technologist on both adult and pediatric subjects taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities and other body parts.
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
GE Healthcare has become aware of a potential safety issue associated with the improper installation of the Precision 600FP monitor suspension. A partial detachment of the monitor suspension could result in potential injury to a patient or operator.
FDA root cause
Other
Action
The firm issued an urgent medical device correction notice by letter on 12/10/2020. The letter states: "You can continue to use your Precision 600 FP monitor suspension under the following condition. If you observe any abnormal movement or misalignment of the monitor suspension in day to day performance controls, discontinue use and contact your GE Healthcare Service Representative immediately." Customers with questions or concerns may contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity in commerce
113 units
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NJ, NM, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, WI, WV
Product codes and lots
Serial Numbers: G1D1842032, G1D1842035, G1C17X2022, G1D1852039, Not Available*, G1D1922084, G1B16X2002, G1D1972106, G1D18X2058, G1D18Y2066, G1B1912074, G1B1772014, G1C1782016, G1D1842033, G1D18Z2071, G1D1962102, G1D1962104, G1D1862041, G1D18Y2068, G1D1972107, G1D18Z2069, Not Available*, G1D19X2118, G1D1982112, Not Available*, G1D1862043, G1D18Y2065, Not Available*, G1D1962101, G1C17X2026, G1D1832029, G1D1962105, G1B1772012, G1D1832030, G1B1772012, G1D1852038, Not Available, G1D1842031, G1C17X2023, G1C1782017, G1D1892054, G1D1932088, G1C17X2024, G1D1882049, G1D1882048, G1D1912077, G1D1932089, GID194092, G1923001923, G1793810017, G1D1882052, G1D1922080, G1D1862045, G1D1882051, G1C1792020, G1D1892055, G1D1932090, G1B1742006, G1D18Z2070, G1D1862042, G1D18X2062, G1D18X2061, G1B1762011, G1B1762010, G1D1942093, G1D1862044, G1D1882050, G1B1912079, G1D1992113, G1B1762009, G1D1852036, G1D1992117, G1D1842034, G1D1922083, G1D1952097, G1D18Z2073, G1D18X2060, G1D1912078, G1D1922085, G1D1892057, Not Available*, G1D1932091, G1D1922081, Not Available*, G1D1912075, G1D18Z2072, G1D1852040, G1D1892056, G1C1792021, G1D1982111, G1D1952100, G1B16X2005, G1D1952096, G1D1962103, G1D1922082, G1D1972108, Not Available*, G1D1932086, G1B16X2004, G1D1852037, G1D18X2059, G1D18Y2064, G1D18Y2063, G1B1942095, G1D1932087, G1D1913076, G1D18Y2067, G1D1872046, G1D1832028, G1D1992114, G1C1792018, G1C17X2025, G1B16X2003  not available* - to be supplied

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1101-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1101-2021Class IITerminatedVoluntary: Firm initiated