Z-0771-2021Class II

SOMATOM Force, Definition AS, Definition Edge, Definition Flash, Drive, Confidence and Edge Plus

Siemens Medical Solutions USA, Inc. / initiated 2020-11-24 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of SOMATOM Force, Definition AS, Definition Edge, Definition Flash, Drive, Confidence and Edge Plus on . FDA classified it as Class II and gives the reason as "System does not trigger a cancel command followed by a reload of the scan with the new parameter(s). This issue may result in a delay in diagnosis and/or need for patient rescan". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0771-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SOMATOM Force, Definition AS, Definition Edge, Definition Flash, Drive, Confidence and Edge Plus
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
System does not trigger a cancel command followed by a reload of the scan with the new parameter(s). This issue may result in a delay in diagnosis and/or need for patient rescan
Root cause tag
software
FDA root cause
Software design
Action
Customer notification letters will be sent starting December 14, 2020 to all affected customers via Customer Advisory Notice CT057/20/S. The technical solution, software versionVB20A_SP3, will be distributed to all affected customers free of charge via CT056/20/S. For questions or technical support, contact your local application specialists or your local service/sales organization at 1-800-888-7436.
Quantity in commerce
3044
Distribution
US Nationwide
Product codes and lots
SOMATOM Force (Model #10742326) SOMATOM Definition As (Model #8098027) SOMATOM Definition Edge (Model #10590000) SOMATOM Definition Flash (Model #10430603) SOMATOM Drive (Model #10431700) SOMATOM Confidence (Model #10590100) SOMATOM Edge Plus (Model #10267000) Siemens SOMATOM CT scanner systems with software syngo.CT VB20A_SP2.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0771-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0771-2021Class IITerminatedFDA Mandated