FUJIFILM Synapse PACS Software: versions 7.0.0 and 7.0.1 -
Fujifilm Medical Systems U.S.A., Inc. began a recall of FUJIFILM Synapse PACS Software: versions 7.0.0 and 7.0.1 - Product Usage: intended for use as a web based application on an off-the shelf PC which meets or... on . FDA classified it as Class II and gives the reason as "The software does not update measurements and calculations in the Clinical Reporting Application (CRA) when the ventricular trace is changed in the study by a different user". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0918-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- FUJIFILM Synapse PACS Software: versions 7.0.0 and 7.0.1 - Product Usage: intended for use as a web based application on an off-the shelf PC which meets or exceeds minimum specifications and is networked with a FUJIFILM Synapse PACS server.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- The software does not update measurements and calculations in the Clinical Reporting Application (CRA) when the ventricular trace is changed in the study by a different user.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- On 12/11/2020 Fujifilm sent consignee notifications via FedEx Priority Overnight stating that there are issues with Synapse PACS Software Version 7.0.000 and 7.0.100. Fujifilm provided a work around until a hotfix is scheduled and completed.
- Quantity in commerce
- 1 unit
- Distribution
- US Nationwide distribution in the state of MD.
- Product codes and lots
- Versions 7.0.0 and 7.0.1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0918-2021 | Class II | Terminated | Voluntary: Firm initiated |