System, X-Ray, Tomography, Computed: Revolution Apex, Revolution CT with Apex Edition
GE Healthcare, LLC began a recall of System, X-Ray, Tomography, Computed: Revolution Apex, Revolution CT with Apex Edition on . FDA classified it as Class II and gives the reason as "GE Healthcare has become aware of a potential issue on Revolution Apex, Revolution CT with Apex Edition, Revolution CT, and Revolution CT ES systems where the table pinch protector could be damaged or missing leading to...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1129-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- System, X-Ray, Tomography, Computed: Revolution Apex, Revolution CT with Apex Edition
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- GE Healthcare has become aware of a potential issue on Revolution Apex, Revolution CT with Apex Edition, Revolution CT, and Revolution CT ES systems where the table pinch protector could be damaged or missing leading to exposed table screws which could cause a laceration due to a sharp edge.
- FDA root cause
- Process change control
- Action
- The manufacturer sent an "Urgent Medical Device Correction" letter (GEHC Ref# 25496) to customers with affected systems on 12/17/2020.
- Quantity in commerce
- 96 units
- Distribution
- Worldwide - US Nationwide Distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NJ, NV, NY, NH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV; In the countries of Algeria, Australia, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Denmark, Estonia, Finland, France, Gabon, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea (Republic Of), Kuwait, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uzbekistan, Viet Nam.
- Product codes and lots
- Model Numbers: 5590000-20: REV2A1900002CN, REV2A2000013CN, REV2A2000025CN, REV2A2000006CN, REV2A2000003CN, REV2A2000015CN, REV2A1900003CN, REV2A2000007CN, REV2A2000022CN, REV2A2000010CN, REV2A2000030CN, REV2A2000029CN, REV2A2000017CN, REV2A2000001CN, REV2A2000002CN, REV2A2000031CN, REV2A2000020CN, REV2A2000024CN, REV2A2000023CN, REV2A2000012CN, REV2A2000016CN, REV2A1900001CN, REV2A2000027CN, REV2A2000028CN, REV2A2000018CN, REV2A1900005CN, REV2A2000009CN, REV2A2000026CN, REV2A2000021CN, REV2A2000011CN 5590000-21: xxx, xxx, xxx* (66 units) *serial number information current as of 01/21/2021.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1129-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1130-2021 | Class II | Terminated | Voluntary: Firm initiated |