Z-0882-2021Class II

A.L.P.S. Clavicle Plating System, Plate, Fixation, Bone, Item Number 233500004 -

Zimmer GmbH / initiated 2020-12-23 / Terminated / root cause: design

Biomet, Inc. began a recall of A.L.P.S. Clavicle Plating System, Plate, Fixation, Bone, Item Number 233500004 - Product Usage: indicated for fixation of fractures, osteotomies and non-unions... on . FDA classified it as Class II and gives the reason as "The instruments may become cold welded together when the drill is inserted through the Soft Tissue Guide. This prevents disassociation of the instruments and potentially leads to an increased chance of fracture of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0882-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
A.L.P.S. Clavicle Plating System, Plate, Fixation, Bone, Item Number 233500004 - Product Usage: indicated for fixation of fractures, osteotomies and non-unions of the clavicle including osteopenic bone.
Product code
HRS Plate, Fixation, Bone
Reason for recall
The instruments may become cold welded together when the drill is inserted through the Soft Tissue Guide. This prevents disassociation of the instruments and potentially leads to an increased chance of fracture of the drill.
Root cause tag
design
FDA root cause
Device Design
Action
On 12/30/2020, Zimmer Biomet issued an Urgent Medical Device Recall to the consumer/user level, including any intermediate wholesale or retail consignees. All distributors will be notified via email. Hospital risk managers and surgeons, as well as distributors with product, will be notified via courier.
Quantity in commerce
174 systems
Distribution
Nationwide distribution.
Product codes and lots
Item: 233500004; Lot: 516349, 545167; UDI: (01)0088030483464(10)516349, (01)00880304834644(11)190208(10)545167

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0882-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0882-2021Class IITerminatedVoluntary: Firm initiated