A.L.P.S. Clavicle Plating System, Plate, Fixation, Bone, Item Number 233500004 -
Biomet, Inc. began a recall of A.L.P.S. Clavicle Plating System, Plate, Fixation, Bone, Item Number 233500004 - Product Usage: indicated for fixation of fractures, osteotomies and non-unions... on . FDA classified it as Class II and gives the reason as "The instruments may become cold welded together when the drill is inserted through the Soft Tissue Guide. This prevents disassociation of the instruments and potentially leads to an increased chance of fracture of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0882-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- A.L.P.S. Clavicle Plating System, Plate, Fixation, Bone, Item Number 233500004 - Product Usage: indicated for fixation of fractures, osteotomies and non-unions of the clavicle including osteopenic bone.
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- The instruments may become cold welded together when the drill is inserted through the Soft Tissue Guide. This prevents disassociation of the instruments and potentially leads to an increased chance of fracture of the drill.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 12/30/2020, Zimmer Biomet issued an Urgent Medical Device Recall to the consumer/user level, including any intermediate wholesale or retail consignees. All distributors will be notified via email. Hospital risk managers and surgeons, as well as distributors with product, will be notified via courier.
- Quantity in commerce
- 174 systems
- Distribution
- Nationwide distribution.
- Product codes and lots
- Item: 233500004; Lot: 516349, 545167; UDI: (01)0088030483464(10)516349, (01)00880304834644(11)190208(10)545167
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0882-2021 | Class II | Terminated | Voluntary: Firm initiated |