Z-0959-2021Class II

PRYTIME MEDICAL REBOA Catheter Introducer Kit, REF: KT1835C, Contents: Needle, 18 GA x 7 CM ECHOGENIC...

Prytime Medical Devices, Inc. / initiated 2020-12-14 / Terminated

Prytime Medical Devices, Inc. began a recall of PRYTIME MEDICAL REBOA Catheter Introducer Kit, REF: KT1835C, Contents: Needle, 18 GA x 7 CM ECHOGENIC, STERILE, 1 EA, SYRINGE, 30CC, STERILE 1 EA, INTRODUCER... on . FDA classified it as Class II and gives the reason as "There is a potential that a catheter convenience set contains a 21G access needle instead of the correct 18G needle. If the 21G needle is used to gain vascular access then the guidewire included in the kit will not pass...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0959-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Prytime Medical Devices, Inc.
Product
PRYTIME MEDICAL REBOA Catheter Introducer Kit, REF: KT1835C, Contents: Needle, 18 GA x 7 CM ECHOGENIC, STERILE, 1 EA, SYRINGE, 30CC, STERILE 1 EA, INTRODUCER SHEATH KIT, 7 FR W/.035 SHORT GUIDEWIRE, STERILE 1 EA., PREFILLED, STERILE FIELD SALINE SYRINGE, 10 ML, 4 EA, CLAMP, 5 FR, STERILE, 1 EA, SCALPEL, #11SS, STERILE, 1 EA, SUTURE, NYLON, 18", 2-0, REVERSE CUTTING, STERILE 1 EA, THREE QUARTER DRAPE, STERILE 1 EA, INSERT, SALINE 1 EA, INSERT, CLAMP 1 EA UDI: (01) 0 0863092 00010 1
Product code
OFC Catheter Introducer Kit
Reason for recall
There is a potential that a catheter convenience set contains a 21G access needle instead of the correct 18G needle. If the 21G needle is used to gain vascular access then the guidewire included in the kit will not pass through the access needle to facilitate insertion of an introducer sheath and could therefore cause procedure delay or injury.
FDA root cause
Mixed-up of materials/components
Action
On 12/14/2020, Prytime Medical sent an "Urgent Medical Device Recall Notification" to affected consignees via FedEx. In addition to informing consignees about the recall, the firm asked consignees to take the following actions: 1) Please immediately review your inventory for the recalled lot specified in this notification. Quarantine and do NOT use this lot. 2) If you have an urgent need for this kit, and dont have an alternative lot available then please ensure that you have a backup 18 gauge access needle readily available at the point of use. Do not use the 21G access needle in the kit. 3) Immediately complete and return the attached Customer Response Form to the contact on the form. a. Please specify the number of affected kits in your possession. b. Prytime will promptly contact you to arrange return and replacement shortly after receiving the completed form. 4) Share this recall notification with any users of the product within your facilities, or anyone else you may have sent the product to, in order to ensure they are also aware of this recall. 5) Report any adverse health consequences experienced with the use of this product to Prytime using the attached form, or by email at [email protected], or by phone at (210) 340-0116. If you have any questions or discover any additional REBOA Convenience Set KT1835C Lot MFG3590 at your facility please contact the Recall Coordinator by phone at (210) 340-0116, or by email at [email protected].
Distribution
US Nationwide distribution including in the states of NJ, NY, CA, TX, MI, MO, CO, MI, VA, DE, TN, MO, MT, AZ, FL, WI, SC, WA, TX, CT, IN, PA, ME, WI, MA, GA, OH, AR, UT, NY OUS: None
Product codes and lots
Lot # MFG 3590

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0959-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0959-2021Class IITerminatedVoluntary: Firm initiated