Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0875-2021Class II

ACCESS SARS-CoV-2 IgG QC SARS-CoV-2 IgG REF C58964 -

Siemens Healthcare Diagnostics Inc. / initiated 2020-12-30 / Terminated / root cause: labeling and UDI

Beckman Coulter Inc. began a recall of ACCESS SARS-CoV-2 IgG QC SARS-CoV-2 IgG REF C58964 - Product Usage: use on the Access Family of Immunoassay Systems only on . FDA classified it as Class II and gives the reason as "The Quality Control (QC) card has a label error for the standard deviation (SD) values for one out of the two levels of QC". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0875-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ACCESS SARS-CoV-2 IgG QC SARS-CoV-2 IgG REF C58964 - Product Usage: use on the Access Family of Immunoassay Systems only.
Product code
QKO Reagent, Coronavirus Serological
Reason for recall
The Quality Control (QC) card has a label error for the standard deviation (SD) values for one out of the two levels of QC.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
On or about 01/04/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification to customer via mail or email informing them that the Recalling Firm has determined that the QC kit lot includes a Quality Control value card with an incorrect QC1 standard deviation (SD) value which may lead to a QC range that is too wide for the control level. Customers are instructed to: 1) Discard the incorrect QC value card included with the affected SARS-CoV-2 IgG quality control lot. 2) Use the Access SARS-CoV-2 IgG QC value card that is included with the Recall Notification letter in place of the discarded value card. 3) Review the QC1 SD value they have implemented and update if appropriate. Evaluate prior QC1 results against the correct SD value. If they detect unsatisfactory QC1 results, evaluate their previous test results. 4) Share this information with their laboratory staff and retain the recall notification as part of their laboratory Quality System documentation. 5) if the product has been forwarded to another laboratory, please provide the Recall Notification to them. 6) Complete and return/respond within 10 days to the Recall Notification: " Electronically, if they received the communication via email. " Manually, complete and return the enclosed Response Form. The Recalling Firm's actions: -They are no longer distributing the affected lot -All future lots will include a correct QC value card. Any questions - contact Customer Support Center at: " From our website: http://www.beckmancoulter.com " By phone: call 1-800-854-3633 in the United States. " Outside the United States contact your local Beckman Coulter representative.
Quantity in commerce
276 units
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, FL, GA, IA, ID, IL, IN, KS,KY, LA, MD, MI, MN, MO, NC, ND, NY, OH, OK, OR, PA, Puerto Rico, SD, TX, WA, and WV. The countries of Andorra, Belgium, Brazil, Bulgaria, Canada*, Croatia, France, Germany, Hong Kong, Italy, Jordan, Mexico, Paraguay, Poland, Portugal, Romania, Spain, and United Kingdom.
Product codes and lots
Catalog Number: C58964 UDI Code: 15099590738662 Lot Number: 922407

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0875-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0875-2021Class IITerminatedVoluntary: Firm initiated