Nucleic acid-based in vitro diagnostic devices for the detection of 2019-novel coronavirus (2019-nCoV) and/or...

Biomeme, Inc. / initiated 2021-01-05 / Terminated / root cause: software

Biomeme, Inc. began a recall of Nucleic acid-based in vitro diagnostic devices for the detection of 2019-novel coronavirus (2019-nCoV) and/or diagnosis of 2019-nCoV infection in human... on . FDA classified it as Class III and gives the reason as "Users cannot complete testing due to a sign-error in the software component that controls the filter movement and results in an instrument failure and assay failure before results are generated". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0940-2021
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Biomeme, Inc.
Product
Nucleic acid-based in vitro diagnostic devices for the detection of 2019-novel coronavirus (2019-nCoV) and/or diagnosis of 2019-nCoV infection in human clinical specimens. The RT-PCR test runs on a Franklin Real-Time PCR Thermocycler Device (Reference/Catalog Number 1000003,1000018) with the Software Controller version 6.4.3. The cycler has a "baseboard" unit with firmware and associated software component that control motorized movements of the optical filter carriage used for PCR detection.
Product code
QJR Reagents, 2019-Novel Coronavirus Nucleic Acid
Reason for recall
Users cannot complete testing due to a sign-error in the software component that controls the filter movement and results in an instrument failure and assay failure before results are generated.
Root cause tag
software
FDA root cause
Software design
Action
On 01/05/2021 Biomeme, Inc., issued an Urgent Device Recall Correction for the latest build of the Franklin Real-Time PCR Thermocycler, which affects some devices built since September 21st, 2020. This software correction is to fix an issue where the device is unable to get past the beginning stages of the test.
Quantity in commerce
147 systems
Distribution
Domestic: FL, TX, IN, KY, OH, NY, CA, PA, MD. Foreign: Canada, Zimbabwe.
Product codes and lots
Device Reference/Catalog Number (Black) - 1000003, (white) - 1000018. Software version 6.4.3  EUA - Emergency Use Authorization

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0940-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0940-2021Class IIITerminatedVoluntary: Firm initiated