Z-1143-2021Class II

Ref: NT821731C, Natus EDS 3 CSF External Drainage System -

Natus Manufacturing Limited / initiated 2021-01-12 / Terminated / root cause: sterility

Natus Manufacturing Limited began a recall of Ref: NT821731C, Natus EDS 3 CSF External Drainage System - Product Usage: indicated for draining cerebrospinal fluid (CFS) from the cerebral ventricles or the... on . FDA classified it as Class II and gives the reason as "The external drainage system failed to meet Pyrogen test requirements for devices contacting cerebrospinal fluid of 2.15 endotoxins". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-1143-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Natus Manufacturing Limited
Product
Ref: NT821731C, Natus EDS 3 CSF External Drainage System - Product Usage: indicated for draining cerebrospinal fluid (CFS) from the cerebral ventricles or the lumbar subarachnoid space as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated.
Product code
JXG Shunt, Central Nervous System And Components
Reason for recall
The external drainage system failed to meet Pyrogen test requirements for devices contacting cerebrospinal fluid of 2.15 endotoxins.
Root cause tag
sterility
FDA root cause
Release of Material/Component prior to receiving test results
Action
On January 12, 2021, Natus sent an "Urgent Field Safety Notice - Recall" notification to all affected consignees via US Mail. In addition to informing consignees about the recall, the notification asked consignees to take the following actions: 1. Please return these affected items at your earliest convenience to the following address: 3150 Pleasant View Road, Middleton, WI 53562 USA 2. Replacement product is available. The Technical Service department will be in contact in relation to the provision of the replacement product. 3. Please complete and return the customer reply form to Natus at the following Email address: [email protected]
Quantity in commerce
7 units
Distribution
US Nationwide distribution in the state of NY.
Product codes and lots
Lot # 118000422298

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1143-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1143-2021Class IITerminatedVoluntary: Firm initiated