Ref: NT821731C, Natus EDS 3 CSF External Drainage System -
Natus Manufacturing Limited began a recall of Ref: NT821731C, Natus EDS 3 CSF External Drainage System - Product Usage: indicated for draining cerebrospinal fluid (CFS) from the cerebral ventricles or the... on . FDA classified it as Class II and gives the reason as "The external drainage system failed to meet Pyrogen test requirements for devices contacting cerebrospinal fluid of 2.15 endotoxins". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1143-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Natus Manufacturing Limited
- Product
- Ref: NT821731C, Natus EDS 3 CSF External Drainage System - Product Usage: indicated for draining cerebrospinal fluid (CFS) from the cerebral ventricles or the lumbar subarachnoid space as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated.
- Product code
- JXG Shunt, Central Nervous System And Components
- Reason for recall
- The external drainage system failed to meet Pyrogen test requirements for devices contacting cerebrospinal fluid of 2.15 endotoxins.
- Root cause tag
- sterility
- FDA root cause
- Release of Material/Component prior to receiving test results
- Action
- On January 12, 2021, Natus sent an "Urgent Field Safety Notice - Recall" notification to all affected consignees via US Mail. In addition to informing consignees about the recall, the notification asked consignees to take the following actions: 1. Please return these affected items at your earliest convenience to the following address: 3150 Pleasant View Road, Middleton, WI 53562 USA 2. Replacement product is available. The Technical Service department will be in contact in relation to the provision of the replacement product. 3. Please complete and return the customer reply form to Natus at the following Email address: [email protected]
- Quantity in commerce
- 7 units
- Distribution
- US Nationwide distribution in the state of NY.
- Product codes and lots
- Lot # 118000422298
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1143-2021 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.