Z-1142-2021Class II

Covidien Parietex hydrophilic anatomical mesh 15 cm x 10 cm polyester left -

Medtronic, Inc. / initiated 2021-01-13 / Terminated / root cause: labeling and UDI

Covidien Llc began a recall of Covidien Parietex hydrophilic anatomical mesh 15 cm x 10 cm polyester left - Product Usage: designed for the repair of inguinal hernias via laparoscopic or... on . FDA classified it as Class II and gives the reason as "incorrect device is contained in the package. The packaging label indicates that the device is for left side use, the product inside the package is for right side use. Use of an opposite side mesh can result in implant...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1142-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Covidien Parietex hydrophilic anatomical mesh 15 cm x 10 cm polyester left - Product Usage: designed for the repair of inguinal hernias via laparoscopic or posterior open procedures Item Code: TECT1510AL
Product code
FTL Mesh, Surgical, Polymeric
Reason for recall
incorrect device is contained in the package. The packaging label indicates that the device is for left side use, the product inside the package is for right side use. Use of an opposite side mesh can result in implant failure and hernia recurrence
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
Medtronic issued Urgent Medical Device Recall letter via Federal Express and/or certified mail beginning January 13, 2021. The letter states reason for recall, health risk and action to take:: " Immediately quarantine and discontinue use of the affected item codes and lots. " Return affected product. " Forward the information to anyone who they have distributed affected Covidien Parietex" Hydrophilic Anatomical Mesh. " Complete the Recalled Product Return Form. Inventory within Medtronic control and customer returned products will be scrapped. Questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Quantity in commerce
129 units WW
Distribution
Worldwide distribution - US Nationwide distribution in the state of MA and the countries of France, French Guiana, Germany, Luxembourg, Spain, United Kingdom.
Product codes and lots
Lot Number: SUH0709M

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1142-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1142-2021Class IITerminatedVoluntary: Firm initiated