Universal Femoral Sleeves -
DePuy Orthopaedics, Inc. began a recall of Universal Femoral Sleeves - Product Usage: used primarily in revision surgeries with the S-ROM Noiles, LPS, SIGMA, and LCS Knee products on . FDA classified it as Class II and gives the reason as "The taper dimensions may be out of specification due to a production issue during the manufacturing of these lots". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1103-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Depuy Orthopaedics Inc.
- Product
- Universal Femoral Sleeves - Product Usage: used primarily in revision surgeries with the S-ROM Noiles, LPS, SIGMA, and LCS Knee products.
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- The taper dimensions may be out of specification due to a production issue during the manufacturing of these lots.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 01/12/2021, DePuy Synthes issued an Urgent Medical Device Recall (removal) via letter due to taper dimensions of certain Universal Femoral Sleeves may be out of specification. The Universal Femoral Sleeves are used primarily in revision surgeries with the S-ROM Noiles, LPS, SIGMA, and LCS Knee products. DePuy is recommending that customers return the recalled devices.
- Quantity in commerce
- 60 sleeves
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AK, CA, FL, IN, MI, MN, MS, NC, NY, OH, PA, TX, VA and the countries of Canada, China, Germany, Puerto Rico, Switzerland.
- Product codes and lots
- Product No.: 129453216, 129453226, 129453236,129453246 Lot No.: J85Y90, J86T23, J86T20, J8790J, J85Y92