Tube Tracheostomy and Tube Cuff, Bivona Cuffless Neonatal/Pediatric FlexTend" Tracheostomy Tube (model...
Smiths Medical ASD Inc. began a recall of Tube Tracheostomy and Tube Cuff, Bivona Cuffless Neonatal/Pediatric FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) V Neck... on . FDA classified it as Class II and gives the reason as "Smiths Medical became aware that three lot numbers of a specific model of Bivona Cuffless FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) may have been incorrectly packaged with a...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1145-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Tube Tracheostomy and Tube Cuff, Bivona Cuffless Neonatal/Pediatric FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) V Neck Flange 4.5mm - Product Usage: intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed for single-patient use up to 5 times.
- Product code
- JOH Tube Tracheostomy And Tube Cuff
- Reason for recall
- Smiths Medical became aware that three lot numbers of a specific model of Bivona Cuffless FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) may have been incorrectly packaged with a straight neck flange instead of a V-neck flange. As a result, there is potential for patient injury if an incorrect device or accessory is selected for use. This use could lead to an internal injury resulting from excessive device contact with tissue, or an external patient injury due to external portions of the device pressing into or rubbing against tissue.
- Root cause tag
- packaging
- FDA root cause
- Under Investigation by firm
- Action
- Distribution of the "URGENT MEDICAL DEVICE RECALL NOTICE" dated 01/29/2021 during the week of January 18, 2021.
- Quantity in commerce
- 548 units
- Distribution
- US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, IL, KS, KY, LA, MA, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, TX, UT, VA, and WA.
- Product codes and lots
- model 60PFS45, lots 3991171 (202 units), 3991172 (117 units), and 3988103 (229 units).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1145-2021 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY