VITROS Chemistry Products VersaTips -
Ortho-Clinical Diagnostics, Inc. began a recall of VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In... on . FDA classified it as Class II and gives the reason as "A deformity affecting a specific lot of VITROS Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1136-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In addition, VersaTips are used to dispense R1 Reagent Fluid into cuvettes in combination with sample fluid for Micro Tips Assays.
- Product code
- DIP Enzyme Immunoassay, Diphenylhydantoin
- Reason for recall
- A deformity affecting a specific lot of VITROS Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays.
- FDA root cause
- Under Investigation by firm
- Action
- On 01/18/2021, Ortho Clinical Diagnostics issued an Urgent Product Correction Notification via e-mail due to a a deformity in a specific lot of VITROS Chemistry Products VersaTips. The affected product is to be no longer distributed and discarded.
- Quantity in commerce
- 4,062 cartons of 1,000 tips (Totaling 4,062,000 tips)
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI and the country of Canada.
- Product codes and lots
- Lot 413420 Model(s): VITROS 250, VITROS 350, VITROS ECi, VITROS 5,1 FS, VITROS 3600, VITROS 4600, VITROS 5600, VITROS XT7600.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1136-2021 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY