Central Screw 6.5x25mm -
Biomet, Inc. began a recall of Central Screw 6.5x25mm - Product Usage: used as a sterile screw that is intended for implantable shoulder prosthesis assembly and/or fixation on . FDA classified it as Class II and gives the reason as "Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive Reverse Shoulder Central Screw, 25 MM Length. It has been determined through a complaint investigation that a 40 MM length central...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1131-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Central Screw 6.5x25mm - Product Usage: used as a sterile screw that is intended for implantable shoulder prosthesis assembly and/or fixation.
- Product code
- PHX Shoulder Prosthesis, Reverse Configuration
- Reason for recall
- Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive Reverse Shoulder Central Screw, 25 MM Length. It has been determined through a complaint investigation that a 40 MM length central screw was packaged and labeled as a 25 MM length central screw.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- On 01/19/2019, Zimmer Biomet issued Urgent Medical Device Recall notices to distributors with direct sales and/or fielded inventory via email and traceable letter sent via FedEx. Hospitals with direct sales or current inventory will be notified via FedEx. Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive Reverse Shoulder Central Screw, 25 MM Length. It has been determined through a complaint investigation that a 40 MM length central screw was packaged and labeled as a 25 MM length central screw.
- Quantity in commerce
- 49
- Distribution
- US Nationwide distribution including in the states of MS, MN, KY, FL, MI, OH, AZ, IL, WI, CA, UT, TN.
- Product codes and lots
- Model No.: 115395, Lot No.: 077170, UDI: (01) 00880304677074 (17) 301010 (10) 077170
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1131-2021 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.