Captivator/Captiflex/Sensation Short Throw Single-Use Polypectomy Snares Outer box UPN, Inner pouch UPN...
Boston Scientific Corporation began a recall of Captivator/Captiflex/Sensation Short Throw Single-Use Polypectomy Snares Outer box UPN, Inner pouch UPN: M00560311 (Box 5), M00560310; M00560321 (Box 5)... on . FDA classified it as Class II and gives the reason as "Packaging Incomplete Seal. Users noted an incomplete pouch seal prior to using the device in a patient". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1170-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Boston Scientific Corporation
- Product
- Captivator/Captiflex/Sensation Short Throw Single-Use Polypectomy Snares Outer box UPN, Inner pouch UPN: M00560311 (Box 5), M00560310; M00560321 (Box 5), M00560320; M00562651 (Box 5), M00562650; M00562652 (Box 20), M00562650; M00562671 (Box 5), M00562670; M00562672 (Box 20), M00562670; M00562673 (Box 40), M00562670; M00562691 (Box 5); M00562690; M00562692 (Box 20), M00562690; M00562693 (Box 40), M00562690.
- Product code
- FDI Snare, Flexible
- Reason for recall
- Packaging Incomplete Seal. Users noted an incomplete pouch seal prior to using the device in a patient.
- Root cause tag
- packaging
- FDA root cause
- Process change control
- Action
- On 01/21/2021, Boston Scientific issued Urgent Medical Device Product Removal notices to customers via e-mail. The removal notice is due to an incomplete seal of the inner pouch which could result in a sterile barrier breach rendering the specified lots of polypectomy snare devices non-sterile.
- Quantity in commerce
- 335,788 units total (Domestic: 222,882, Foreign: 112,906)
- Distribution
- US nationwide distribution.
- Product codes and lots
- Outer box GTIN: 08714729283904 Inner pouch GTIN: 08714729747598 Lot/Batch#: 24818278, 24906823, 25042455 Outer box GTIN: 08714729283928 Inner pouch GTIN: 08714729747611 Lot/Batch#: 24685766, 24817145, 24899710, 24739786, 25006735, 24861901 Outer box GTIN: 08714729158110 Inner pouch GTIN: 08714729748069 Lot/Batch#: 24685764, 24890051, 24824132 Outer box GTIN: 08714729268802 Inner pouch GTIN: 08714729748069 Lot/Batch#: 25023345 Outer box GTIN: 08714729158141 Inner pouch GTIN: 08714729748083 Lot/Batch#: 24699718, 24939626, 24861739 Outer box GTIN: 08714729268819 Inner pouch GTIN: 08714729748083 Lot/Batch#: 24824134, 25006729, 24709534 Outer box GTIN: 08714729501664 Inner pouch GTIN: 08714729748083 Lot/Batch#: 25023203, 24697822, 24739587, 24942292, 24909214, 24824136 Outer box GTIN: 08714729158158 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24856636, 24927642, 24699218 Outer box GTIN: 08714729268826 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24835709, 24818605, 25023343 Outer box GTIN: 08714729501671 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24909216, 24758832, 24872119, 24994171, 25048397, 2475126, 24929431, 24995438
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1168-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1169-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1170-2021 | Class II | Ongoing | Voluntary: Firm initiated |