Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1170-2021Class II

Captivator/Captiflex/Sensation Short Throw Single-Use Polypectomy Snares Outer box UPN, Inner pouch UPN...

Boston Scientific Corporation / initiated 2021-01-19 / Open, Classified / root cause: packaging

Boston Scientific Corporation began a recall of Captivator/Captiflex/Sensation Short Throw Single-Use Polypectomy Snares Outer box UPN, Inner pouch UPN: M00560311 (Box 5), M00560310; M00560321 (Box 5)... on . FDA classified it as Class II and gives the reason as "Packaging Incomplete Seal. Users noted an incomplete pouch seal prior to using the device in a patient". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1170-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Boston Scientific Corporation
Product
Captivator/Captiflex/Sensation Short Throw Single-Use Polypectomy Snares Outer box UPN, Inner pouch UPN: M00560311 (Box 5), M00560310; M00560321 (Box 5), M00560320; M00562651 (Box 5), M00562650; M00562652 (Box 20), M00562650; M00562671 (Box 5), M00562670; M00562672 (Box 20), M00562670; M00562673 (Box 40), M00562670; M00562691 (Box 5); M00562690; M00562692 (Box 20), M00562690; M00562693 (Box 40), M00562690.
Product code
FDI Snare, Flexible
Reason for recall
Packaging Incomplete Seal. Users noted an incomplete pouch seal prior to using the device in a patient.
Root cause tag
packaging
FDA root cause
Process change control
Action
On 01/21/2021, Boston Scientific issued Urgent Medical Device Product Removal notices to customers via e-mail. The removal notice is due to an incomplete seal of the inner pouch which could result in a sterile barrier breach rendering the specified lots of polypectomy snare devices non-sterile.
Quantity in commerce
335,788 units total (Domestic: 222,882, Foreign: 112,906)
Distribution
US nationwide distribution.
Product codes and lots
Outer box GTIN: 08714729283904 Inner pouch GTIN: 08714729747598 Lot/Batch#: 24818278, 24906823, 25042455  Outer box GTIN: 08714729283928 Inner pouch GTIN: 08714729747611 Lot/Batch#: 24685766, 24817145, 24899710, 24739786, 25006735, 24861901  Outer box GTIN: 08714729158110 Inner pouch GTIN: 08714729748069 Lot/Batch#: 24685764, 24890051, 24824132  Outer box GTIN: 08714729268802 Inner pouch GTIN: 08714729748069 Lot/Batch#: 25023345  Outer box GTIN: 08714729158141 Inner pouch GTIN: 08714729748083 Lot/Batch#: 24699718, 24939626, 24861739  Outer box GTIN: 08714729268819 Inner pouch GTIN: 08714729748083 Lot/Batch#: 24824134, 25006729, 24709534  Outer box GTIN: 08714729501664 Inner pouch GTIN: 08714729748083 Lot/Batch#: 25023203, 24697822, 24739587, 24942292, 24909214, 24824136  Outer box GTIN: 08714729158158 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24856636, 24927642, 24699218  Outer box GTIN: 08714729268826 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24835709, 24818605, 25023343  Outer box GTIN: 08714729501671 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24909216, 24758832, 24872119, 24994171, 25048397, 2475126, 24929431, 24995438

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1170-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1168-2021Class IIOngoingVoluntary: Firm initiated
Z-1169-2021Class IIOngoingVoluntary: Firm initiated
Z-1170-2021Class IIOngoingVoluntary: Firm initiated