Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1152-2021Class II

Milex Gellhorn Flexible under the following labels: 1) Gellhorn Flexible w/Short Stem Pessary, REF...

CooperSurgical, Inc. / initiated 2021-01-21 / Terminated / root cause: manufacturing process

CooperSurgical, Inc. began a recall of Milex Gellhorn Flexible under the following labels: 1) Gellhorn Flexible w/Short Stem Pessary, REF MXKPGSS2-1/4, 2-1/4" (57 mm) with Kit, 2) Gellhorn Flexible... on . FDA classified it as Class II and gives the reason as "The affected Milex Gellhorn pessaries were incorrectly manufactured with a long stem (2.170 ), instead of a short stem (1.670 or 1.960 nominal height) as required. The length discrepancy does not meet the released print...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1152-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CooperSurgical, Inc.
Product
Milex Gellhorn Flexible under the following labels: 1) Gellhorn Flexible w/Short Stem Pessary, REF MXKPGSS2-1/4, 2-1/4" (57 mm) with Kit, 2) Gellhorn Flexible w/Short Stem Pessary, REF MXPGSS2-1/4, 2-1/4" (57 mm) without Kit, 3) Gellhorn Flexible w/Short Stem Pessary, REF MXPGSS2-3/4, 2-3/4" (70 mm)
Product code
HHW Pessary, Vaginal
Reason for recall
The affected Milex Gellhorn pessaries were incorrectly manufactured with a long stem (2.170 ), instead of a short stem (1.670 or 1.960 nominal height) as required. The length discrepancy does not meet the released print dimensions.
Root cause tag
manufacturing process
FDA root cause
Process change control
Action
On 01/21/2021, CooperSurgical issued Urgent Medical Device Recall notices to customers via certified mail. On 9/08/2021, CooperSurgical expanded the recall adding an additional lot and notified customers via courier service. Customers were advised to: - Discontinue use of and isolate all products subject to the recall. - Complete the Acknowledgement and Receipt Form, and return via email to [email protected] or fax to 203-601-9870. - Once the Acknowledgement Form is received, Cooper Surgical will arrange for Product replacement(s) and return of affected Product(s). The replacement Product(s) will be charged at the same price as the original order and your account will be credited back once the affected Product(s) are returned to CooperSurgical. If you do not have affected Product in inventory, please use the same Form to indicate that and return it to CooperSurgical. Customers with questions, please call 203-601-5200 ext. 3300.
Quantity in commerce
109 units (Domestic: 60 units, Foreign: 49 units)
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AZ, CA, FL, IL, KS, MA, MI, NC, NY, PA, TN, TX, WA, & Puerto Rico. The countries of New Zealand, United Kingdom.
Product codes and lots
Recall letter dated 1/21/2021: 1) With Kit: REF #: MXKPGSS2-1/4, Lot #: 276529- Exp Date June 10, 2025, 2) Without Kit: REF #: MXPGSS2-1/4, Lot #: 283083- Exp Date May 12, 2025; Lot 288491- Exp Date May 14, 2025; Lot # 288492- Exp Date May 29, 2025. Recall letter dated 9/8/2021 included one additional lot: 3) REF # MXPGSS2-3/4, Lot 276637- Exp Date February 12, 2026

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1152-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1152-2021Class IITerminatedVoluntary: Firm initiated