OT1000 Series Orthopedic Surgical Tables
Steris Corporation began a recall of OT1000 Series Orthopedic Surgical Tables on . FDA classified it as Class II and gives the reason as "On January 7, 2021, STERIS identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1149-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- STERIS Corporation
- Product
- OT1000 Series Orthopedic Surgical Tables
- Product code
- JEA Table, Surgical With Orthopedic Accessories, Ac-Powered
- Reason for recall
- On January 7, 2021, STERIS identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- In January, 2021, Steris will issue an Urgent Medical Device Correction notice to customers via letter for the OT1000 Series Orthopedic Surgical Table. Steris has identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.
- Quantity in commerce
- 272 tables total (Domestic: 227 tables, Foreign 45 tables)
- Distribution
- Domestic: AL, AK, AR, AZ, CA, CO, CT, DC, FL, GA, ID, IL IN, IW, KS, KY, LA, ME, MD, MI, MN, MO, NE, NV, NJ, NY, NC, ND, OH, OK, OR, PA, PR, SC, TX, UT, VT, VA, WA, WV, WI. Foreign: Canada, South Korea, Panama, Qatar, Thailand.
- Product codes and lots
- Model No.: OT1000, OT1100, OT1200
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1149-2021 | Class II | Terminated | Voluntary: Firm initiated |